Feasibility of adenosine stress cardiovascular magnetic resonance perfusion imaging in patients with MR-conditional transvenous permanent pacemakers and defibrillators

J Cardiovasc Magn Reson. 2022 Jan 13;24(1):9. doi: 10.1186/s12968-021-00842-0.

Abstract

Background: The use of stress perfusion-cardiovascular magnetic resonance (CMR) imaging remains limited in patients with implantable devices. The primary goal of the study was to assess the safety, image quality, and the diagnostic value of stress perfusion-CMR in patients with MR-conditional transvenous permanent pacemakers (PPM) or implantable cardioverter-defibrillators (ICD).

Methods: Consecutive patients with a transvenous PPM or ICD referred for adenosine stress-CMR were enrolled in this single-center longitudinal study. The CMR protocol was performed using a 1.5 T system according to current guidelines while all devices were put in MR-mode. Quality of cine, late-gadolinium-enhancement (LGE), and stress perfusion sequences were assessed. An ischemia burden of ≥ 1.5 segments was considered significant. We assessed the safety, image quality and the occurrence of interference of the magnetic field with the implantable device. In case of ischemia, we also assessed the correlation with the presence of significant coronary lesions on coronary angiography.

Results: Among 3743 perfusion-CMR examinations, 66 patients had implantable devices (1.7%). Image quality proved diagnostic in 98% of cases. No device damage or malfunction was reported immediately and at 1 year. Fifty patients were continuously paced during CMR. Heart rate and systolic blood pressure remained unchanged during adenosine stress, while diastolic blood pressure decreased (p = 0.007). Six patients (9%) had an ischemia-positive stress CMR and significant coronary stenoses were confirmed by coronary angiography in all cases.

Conclusion: Stress perfusion-CMR is safe, allows reliable ischemia detection, and provides good diagnostic value.

Keywords: Adenosine; Cardiovascular magnetic resonance; Implantable cardioverter defibrillator; Implantable device; MRI conditional; Pacemaker; Safety; Stress test.

MeSH terms

  • Adenosine
  • Defibrillators, Implantable*
  • Feasibility Studies
  • Humans
  • Longitudinal Studies
  • Magnetic Resonance Imaging
  • Magnetic Resonance Imaging, Cine
  • Magnetic Resonance Spectroscopy
  • Myocardial Perfusion Imaging*
  • Pacemaker, Artificial*
  • Perfusion Imaging
  • Predictive Value of Tests

Substances

  • Adenosine