A 10 J shock impedance in sinus rhythm correlates with a 65 J defibrillation impedance during subcutaneous defibrillator implantation using an intermuscular technique

J Cardiovasc Electrophysiol. 2021 Nov;32(11):3027-3034. doi: 10.1111/jce.15249. Epub 2021 Sep 28.

Abstract

Introduction: Defibrillation testing (DT) is recommended during the subcutaneous defibrillator (S-ICD) placement. We sought to compare 10 J shock impedance in sinus rhythm (SR) with 65 J defibrillation impedance and evaluate device position on a postimplant chest X-ray (CXR) using an intermuscular (IM) technique.

Methods: Consecutive S-ICD implantations between 12/2019 and 12/2020 at The Ohio State University were reviewed. All implantations were performed using a two-incision IM technique. Standard DT with 65 J shock and 10 J shock in SR were performed unless contraindicated. The PRAETORIAN score was calculated based on CXR.

Results: A total of 37 patients (age: 47.2 ± 15.8 years old, male: n = 26 [70.3%], body mass index: 30.1 ± 6.7 kg/m2 ) underwent IM S-ICD implantation, and of those, 27 (73%) underwent both 65 J shock and 10 J shock in SR. The coefficient of determination (R2 ) between 10 J shock impedance and 65 J shock impedance was 0.84. The mean of an impedance difference was 1.6 ± 4.8 Ω (minimum - 11 and maximum 8). Postimplant CXR was available for 33 out of 37 patients (89.2%). The PRAETORIAN score was less than 90 in all patients and the mean score was 32.7 ± 8.8.

Conclusion: We demonstrated that 10 J shock impedance in SR correlated well with 65 J defibrillation impedance during IM S-ICD implantation. An IM implantation technique provides excellent generator location on postimplant CXR. The IM technique combined with 10 J shock in SR may be sufficient to predict and ensure the defibrillation efficacy of the S-ICD.

Keywords: anesthesia; defibrillation testing; impedance; intermuscular; subcutaneous defibrillator.

MeSH terms

  • Adult
  • Arrhythmias, Cardiac
  • Defibrillators, Implantable*
  • Electric Countershock
  • Electric Impedance
  • Humans
  • Male
  • Middle Aged
  • Ohio
  • Subcutaneous Tissue