Cardiac Magnetic Resonance Assessment of Response to Cardiac Resynchronization Therapy and Programming Strategies

JACC Cardiovasc Imaging. 2021 Dec;14(12):2369-2383. doi: 10.1016/j.jcmg.2021.06.015. Epub 2021 Aug 18.

Abstract

Objectives: The objective was to determine the feasibility and effectiveness of cardiac magnetic resonance (CMR) cine and strain imaging before and after cardiac resynchronization therapy (CRT) for assessment of response and the optimal resynchronization pacing strategy.

Background: CMR with cardiac implantable electronic devices can safely provide high-quality right ventricular/left ventricular (LV) ejection fraction (RVEF/LVEF) assessments and strain.

Methods: CMR with cine imaging, displacement encoding with stimulated echoes for the circumferential uniformity ratio estimate with singular value decomposition (CURE-SVD) dyssynchrony parameter, and scar assessment was performed before and after CRT. Whereas the pre-CRT scan constituted a single "imaging set" with complete volumetric, strain, and scar imaging, multiple imaging sets with complete strain and volumetric data were obtained during the post-CRT scan for biventricular pacing (BIVP), LV pacing (LVP), and asynchronous atrial pacing modes by reprogramming the device outside the scanner between imaging sets.

Results: 100 CMRs with a total of 162 imaging sets were performed in 50 patients (median age 70 years [IQR: 50-86 years]; 48% female). Reduction in LV end-diastolic volumes (P = 0.002) independent of CRT pacing were more prominent than corresponding reductions in right ventricular end-diastolic volumes (P = 0.16). A clear dependence of the optimal CRT pacing mode (BIVP vs LVP) on the PR interval (P = 0.0006) was demonstrated. The LVEF and RVEF improved more with BIVP than LVP with PR intervals ≥240 milliseconds (P = 0.025 and P = 0.002, respectively); the optimal mode (BIVP vs LVP) was variable with PR intervals <240 milliseconds. A lower pre-CRT displacement encoding with stimulated echoes (DENSE) CURE-SVD was associated with greater improvements in the post-CRT CURE-SVD (r = -0.69; P < 0.001), LV end-systolic volume (r = -0.58; P < 0.001), and LVEF (r = -0.52; P < 0.001).

Conclusions: CMR evaluation with assessment of multiple pacing modes during a single scan after CRT is feasible and provides useful information for patient care with respect to response and the optimal pacing strategy.

Keywords: cardiac magnetic resonance; cardiac resynchronization therapy; heart failure; implantable cardioverter-defibrillator.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy* / methods
  • Female
  • Heart Failure* / diagnostic imaging
  • Heart Failure* / therapy
  • Humans
  • Magnetic Resonance Spectroscopy
  • Male
  • Predictive Value of Tests
  • Stroke Volume
  • Treatment Outcome
  • Ventricular Function, Left