Clinical implications of compromised 82Rb PET data acquisition

J Nucl Cardiol. 2022 Oct;29(5):2583-2594. doi: 10.1007/s12350-021-02774-9. Epub 2021 Aug 20.

Abstract

Background: We wished to document the prevalence and quantitative effects of compromised 82Rb PET data acquisitions on myocardial flow reserve (MFR).

Methods and results: Data were analyzed retrospectively for 246 rest and regadenoson-stress studies of 123 patients evaluated for known or suspected CAD. An automated injector delivered pre-determined activities of 82Rb. Automated quality assurance algorithms identified technical problems for 7% (9/123) of patients. Stress data exhibited 2 instances of scanner saturation, 1 blood peak detection, 1 blood peak width, 1 gradual patient motion, and 2 abrupt patient motion problems. Rest data showed 1 instance of blood peak width and 2 abrupt patient motion problems. MFR was lower for patients with technical problems flagged by the quality assurance algorithms than those without technical problems (1.5 ± 0.5 versus 2.1 ± 0.7, P = 0.01), even though rest and stress ejection fraction, asynchrony and relative myocardial perfusion measures were similar for these two groups (P > 0.05), suggesting that MFR accuracy was adversely affected by technical errors.

Conclusion: It is important to verify integrity of 82Rb data to ensure MFR computation quality.

Keywords: 82Rb; Myocardial blood flow; PET; Quality assurance.