The Allegra transcatheter heart valve: Short term results from a multicenter registry

Catheter Cardiovasc Interv. 2021 Nov 15;98(6):1204-1209. doi: 10.1002/ccd.29833. Epub 2021 Jun 17.

Abstract

Objectives: We aimed to determine the safety and efficacy of the Allegra transcatheter heart valve (THV) for the treatment of severe aortic valve stenosis in a large patient population treated under real-world conditions.

Background: The Allegra is a novel self-expanding THV with supra-annular bovine leaflets. The valve is available in three different sizes (23, 27, and 31 mm), all are delivered through an 18F sheath.

Methods: Consecutive patients undergoing TAVR with the Allegra THV were enrolled in a multicenter-registry. Data were collected throughout initial hospital-stay and at 30-day follow-up. Clinical endpoints were defined according to the updated definitions of the Valve-Academic-Research-Consortium.

Results: This registry included 255 patients (mean age 83 ± 6 years, 48% women) from four European centers. Median European System for Cardiac Operative Risk Evaluation II score (EuroSCORE II) was 3.3% (IQR 1.9-5.8%). Acute device success was 95.7%. The remaining 11 patients had either moderate paravalvular regurgitation immediately after the procedure (7 patients) or the device could not be optimal positioned requiring implantation of a second THV (4 patients). Major vascular complications and major/life-threatening bleedings occurred in 10 (3.9%) and 12 (4.7%) patients, respectively. At 30 day follow-up, mean effective orifice area was 2.2 ± 0.5 cm2 , mean gradient was 6.9 ± 3.8 mmHg, 7 (3.3%) patients had more than mild paravalvular leakage, 3 patients (1.2%) had died, 6 patients (2.4%) had a stroke and 30 (12.8%) patients had required implantation of a new permanent pacemaker.

Conclusions: Transfemoral implantation of the Allegra THV resulted in favorable clinical and echocardiographic outcomes during hospitalization and short-term follow up.

Keywords: Allegra; TAVR; aortic stenosis; outcomes; transcatheter aortic valve; transcatheter aortic valve replacement.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Animals
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Cattle
  • Female
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Prosthesis Design
  • Registries
  • Risk Factors
  • Terfenadine / analogs & derivatives
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Treatment Outcome

Substances

  • Terfenadine
  • fexofenadine