Predictors and Clinical Impact of Prosthesis-Patient Mismatch After Self-Expandable TAVR in Small Annuli

JACC Cardiovasc Interv. 2021 Jun 14;14(11):1218-1228. doi: 10.1016/j.jcin.2021.03.060.

Abstract

Objectives: The aim of this study was to define predictors of prosthesis-patient mismatch (PPM) and its impact on mortality after transcatheter aortic valve replacement (TAVR) with self-expandable valves (SEVs) in patients with small annuli.

Background: TAVR seems to reduce the risk for PPM compared with surgical aortic valve replacement, especially in patients with small aortic annuli. Nevertheless, predictors and impact of PPM in this population have not been clarified yet.

Methods: Predictors of PPM and all-cause mortality were investigated using multivariable logistic regression analysis from the cohort of the TAVI-SMALL (International Multicenter Registry to Evaluate the Performance of Self-Expandable Valves in Small Aortic Annuli) registry, which included patients with severe aortic stenosis and small annuli (annular perimeter <72 mm or area <400 mm2 on computed tomography) treated with transcatheter SEVs: 445 patients with (n = 129) and without (n = 316) PPM were enrolled.

Results: Intra-annular valves conferred increased risk for PPM (odds ratio [OR]: 2.36; 95% confidence interval [CI]: 1.16 to 4.81), while post-dilation (OR: 0.46; 95% CI: 0.25-0.84) and valve oversizing (OR: 0.53; 95% CI: 0.28-1.00) seemed to protect against PPM occurrence. At a median follow-up of 354 days, patients with severe PPM, but not those with moderate PPM, had a higher all-cause mortality rate compared with those without PPM (log-rank p = 0.008). Multivariable Cox regression confirmed severe PPM as an independent predictor of all-cause mortality (hazard ratio: 4.27; 95% CI: 1.34 to 13.6).

Conclusions: Among patients with aortic stenosis and small aortic annuli undergoing transcatheter SEV implantation, use of intra-annular valves yielded higher risk for PPM; conversely, post-dilation and valve oversizing protected against PPM occurrence. Severe PPM was independently associated with all-cause mortality.

Keywords: TAVR; prosthesis-patient mismatch; self-expandable valves; small annuli.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Humans
  • Prosthesis Design
  • Risk Factors
  • Transcatheter Aortic Valve Replacement* / adverse effects