Transcatheter Aortic Valve Replacement for Degenerated Transcatheter Aortic Valves: The TRANSIT International Project

Circ Cardiovasc Interv. 2021 Jun;14(6):e010440. doi: 10.1161/CIRCINTERVENTIONS.120.010440. Epub 2021 Jun 7.

Abstract

Background: Transcatheter aortic valve replacement (TAVR) has determined a paradigm shift in the treatment of patients with severe aortic stenosis. However, the durability of bioprostheses is still a matter of concern, and little is known about the management of degenerated TAV. We sought to evaluate the outcomes of patients with a degenerated TAV treated by means of a second TAVR.

Methods: The TRANSIT is an international registry that included cases of degenerated TAVR from 28 centers. Among around 40 000 patients treated with TAVR in the participating centers, 172 underwent a second TAVR: 57 (33%) for a mainly stenotic degenerated TAV, 97 (56%) for a mainly regurgitant TAV, and 18 (11%) for a combined degeneration. Overall, the rate of New York Heart Association class III/IV at presentation was 73.5%.

Results: Valve Academic Research Consortium 2 device success rate was 79%, as a consequence of residual gradient (14%) or regurgitation (7%). At 1 month, the overall mortality rate was 2.9%, while rates of new hospitalization and New York Heart Association class III/IV were 3.6% and 7%, respectively, without significant difference across the groups. At 1 year, the overall mortality rate was 10%, while rates of new hospitalization and New York Heart Association class III/IV were 7.6% and 5.8%, respectively, without significant difference across the groups. No cases of valve thrombosis were recorded.

Conclusions: Selected patients with a degenerated TAV may be safely and successfully treated by means of a second TAVR. This finding is of crucial importance for the adoption of the TAVR technology in a lower risk and younger population.

Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT04500964.

Keywords: bioprosthesis; heart failure; transcatheter aortic valve replacement.

MeSH terms

  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Bioprosthesis*
  • Heart Valve Prosthesis Implantation*
  • Heart Valve Prosthesis*
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT04500964