Gain in real-world cardiac resynchronization therapy efficacy with SyncAV dynamic optimization: Heart failure hospitalizations and costs

Heart Rhythm. 2021 Sep;18(9):1577-1585. doi: 10.1016/j.hrthm.2021.05.006. Epub 2021 May 7.

Abstract

Background: SyncAV, a device-based cardiac resynchronization therapy (CRT) algorithm, promotes electrical optimization by dynamically adjusting atrioventricular intervals.

Objective: The purpose of this study was to evaluate the impact of SyncAV on heart failure hospitalizations (HFHs) and related costs in a real-world CRT cohort.

Methods: Patients with SyncAV-capable CRT devices followed by remote monitoring and enrolled in Medicare fee-for-service for at least 1 year preimplant and up to 2 years postimplant were studied. Patients with SyncAV OFF were 4:1 matched to those with SyncAV ON on preimplant HFH rate, demographics, comorbidities, disease etiology, and left bundle branch block. HFHs were determined from the primary diagnosis of inpatient hospitalizations, and the cost for each event was the sum of Medicare, supplemental insurance, and patient payment.

Results: After 4:1 propensity score matching, 3630 patients were studied (mean age 75 ± 8 years; 1386 [38%] female), including 726 (25%) patients with SyncAV ON. The pre-CRT HFH rate was 0.338 HFH events per patient-year. Overall, CRT diminished the HFH rate to 0.204 events per patient-year (P < .001). SyncAV elicited a larger reduction in HFH rate (SyncAV ON: hazard ratio [HR] 0.52; 95% confidence interval [CI] 0.41-0.66; P < .001 and SyncAV OFF: HR 0.68; 95% CI 0.59-0.77; P < .001). After 2 years, the HFH rate was lower in the SyncAV ON group than in the SyncAV OFF group (0.143 HFHs per patient-year vs 0.193 HFHs per patient-year; HR 0.70; 95% CI 0.55-0.89; P = .003) and fewer HFHs were followed by 30-day HFH readmissions (4.41% vs 7.68%; P = .003) and 30-day all-cause hospital readmissions (7.04% vs 10.01%; P = .010). The total 2-year HFH-associated costs per patient were lower with SyncAV ON (difference $1135; 90% CI $93-$2109; P = .038).

Conclusion: This large, real-world, propensity score-matched study demonstrates that SyncAV CRT is associated with significantly reduced HFHs and associated costs, incremental to standard CRT.

Keywords: Cardiac resynchronization therapy; Heart failure hospitalization; Optimization; Propensity score matched; Readmissions; SyncAV.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Bundle-Branch Block / epidemiology
  • Bundle-Branch Block / therapy
  • Cardiac Resynchronization Therapy Devices* / economics
  • Cardiac Resynchronization Therapy Devices* / standards
  • Cardiac Resynchronization Therapy* / economics
  • Cardiac Resynchronization Therapy* / methods
  • Cardiac Resynchronization Therapy* / statistics & numerical data
  • Comorbidity
  • Cost-Benefit Analysis
  • Fee-for-Service Plans
  • Female
  • Heart Failure / diagnosis
  • Heart Failure / epidemiology
  • Heart Failure / therapy
  • Hospitalization / economics
  • Hospitalization / statistics & numerical data
  • Humans
  • Male
  • Medicare / statistics & numerical data
  • Middle Aged
  • Monitoring, Ambulatory / methods
  • Monitoring, Ambulatory / statistics & numerical data
  • Propensity Score
  • Quality Improvement
  • Treatment Outcome
  • United States