Objectives: This study sought to obtain in vivo data on a new stent and delivery system specifically designed for implantation in infants with the ability to be enlarged to adult dimensions.
Background: There are no endovascular stents designed for or approved for use in infants, nor is there a stent capable of being implanted at infant vessel diameters and achieving adult size while maintaining structural integrity. The Minima stent was designed to address these needs.
Methods: This study was performed in 6 piglets who underwent implantation of 22 Minima stents into the following locations: aorta (n = 11), branch pulmonary arteries (n = 6), and central veins (n = 5).
Results: Successful deployment occurred in 21/22 attempts. Two instances of post-deployment migration occurred. Stents were re-expanded at 1, 2, 3 and 5 months after implant. All stents regardless of location could be re-dilated to the intended diameter to keep pace with somatic growth (implant diameter 6.9 +/- 1.2 mm; final diameter 16.1 mm +/- 1.4 mm). Histopathology at 1 and 5 months demonstrated widely patent vessel lumens with stent apposition to vessel wall, early mild inflammatory response surrounding stent struts, typical vascular damage and healing response to acute dilation and a progressive smooth neointimal growth covering stent struts over time.
Conclusions: In this in vivo study of the Minima stent, there was high implant success, predictable re-dilatability to adult diameters and favorable histopathology. Further study is warranted.
Keywords: congenital heart disease; pediatric intervention; stent; vascular stenosis.
© 2021 The Authors. Catheterization and Cardiovascular Interventions published by Wiley Periodicals LLC.