Scheduled versus alert transmissions for remote follow-up of cardiac implantable electronic devices: Clinical relevance and resource consumption

Int J Cardiol. 2021 Jul 1:334:49-54. doi: 10.1016/j.ijcard.2021.04.043. Epub 2021 Apr 27.

Abstract

Background: The remote follow-up of pacemakers and implantable cardiac defibrillators (ICDs) usually includes scheduled checks and alert transmissions. However, this results in a high volume of remote data reviews to be managed. We measured the relative contribution of scheduled and alert transmissions to the detection of relevant conditions, and the workload generated by their management.

Methods: At our center, the frequency of remote scheduled transmissions is 4/year. Moreover, all system-integrity and clinical alerts are turned on for wireless notification. We calculated the number of transmissions received from January to December 2020, and identified transmissions that necessitated in-hospital access for further assessment and transmissions that required clinical discussion with the physician. For all alert transmissions, we identified whether the alert was clinically meaningful (i.e. center was not previously aware of the condition and no action had yet been taken to treat it).

Results: Of 8545 transmissions received from 1697 pacemakers and ICDs, 5766 (67%) were scheduled and 2779 (33%) were alert transmissions received from 764 patients (45%); 499 (9%) scheduled transmissions required clinical discussion with the physician, but only 2 of these necessitated in-hospital visits for further assessment. Of the alert transmissions, 664 (24%) required clinical discussion, and 75 (3%) necessitated in-hospital visits. The proportion of alerts judged clinically meaningful was 7%.

Conclusion: Scheduled transmissions generate 67% of remote data reviews for pacemakers and ICDs, but their ability to detect clinically relevant events is very low. A strategy that relies exclusively on alert transmissions could ensure continuity of patient monitoring while reducing the workload at the center.

Keywords: ICD; Pacemaker; Remote monitoring; Telecardiology.

MeSH terms

  • Defibrillators, Implantable*
  • Electronics
  • Follow-Up Studies
  • Humans
  • Monitoring, Physiologic
  • Workload