Biodegradable- Versus Durable-Polymer Drug-Eluting Stents for STEMI: Final 2-Year Outcomes of the BIOSTEMI Trial

JACC Cardiovasc Interv. 2021 Mar 22;14(6):639-648. doi: 10.1016/j.jcin.2020.12.011.

Abstract

Objectives: The aim of this study was to investigate the safety and efficacy of biodegradable-polymer sirolimus-eluting stents (BP-SES) compared with durable-polymer everolimus-eluting stents (DP-EES) in patients with ST-segment elevation myocardial infarction (STEMI).

Background: Primary percutaneous coronary intervention (PCI) is an effective treatment for patients with STEMI, and long-term outcomes are determined by the safety and efficacy profile of the newest generation drug-eluting stents.

Methods: BIOSTEMI (A Comparison of an Ultrathin Strut Biodegradable Polymer Sirolimus-Eluting Stent With a Durable Polymer Everolimus-Eluting Stent for Patients With Acute ST-Segment Elevation Myocardial Infarction Undergoing Primary Percutaneous Coronary Intervention) was an investigator-initiated, multicenter, assessor-blind, randomized superiority trial using Bayesian methods. Patients with STEMI undergoing primary PCI within 24 h of symptom onset were randomized in a 1:1 ratio to receive BP-SES (n = 649) or DP-EES (n = 651). The primary endpoint was target lesion failure (TLF), a composite of cardiac death, target vessel myocardial reinfarction, and clinically indicated target lesion revascularization (TLR) at 2 years.

Results: Between April 2016 and March 2018, 1,300 patients were included. Baseline characteristics were comparable between the 2 treatment groups. Follow-up through 2 years was complete in 1,221 patients (94%). At 2 years, TLF occurred in 33 patients (5.1%) treated with BP-SES and in 53 patients (8.1%) treated with DP-EES (rate ratio: 0.58; 95% Bayesian credible interval: 0.40 to 0.84; posterior probability of superiority = 0.998). The difference was driven by a lower incidence of clinically indicated TLR in patients treated with BP-SES compared with DP-EES (2.5% vs. 5.1%; rate ratio: 0.52; 95% Bayesian credible interval: 0.30 to 0.87; posterior probability of superiority = 0.993). There were no significant differences in rates of cardiac death, target vessel myocardial reinfarction, and definite stent thrombosis between the 2 treatment arms.

Conclusions: In patients with STEMI undergoing primary PCI, BP-SES were superior to DP-EES with respect to TLF at 2 years. The difference was driven by lower rates of ischemia-driven TLR. (A Comparison of an Ultrathin Strut Biodegradable Polymer Sirolimus-Eluting Stent With a Durable Polymer Everolimus-Eluting Stent for Patients With Acute ST-Segment Elevation Myocardial Infarction Undergoing Primary Percutaneous Coronary Intervention [BIOSTEMI]; NCT02579031).

Keywords: acute myocardial infarction; biodegradable polymer; drug-eluting stent(s); thin strut.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants
  • Bayes Theorem
  • Drug-Eluting Stents*
  • Humans
  • Percutaneous Coronary Intervention* / adverse effects
  • Polymers
  • Prosthesis Design
  • ST Elevation Myocardial Infarction* / diagnostic imaging
  • ST Elevation Myocardial Infarction* / therapy
  • Treatment Outcome

Substances

  • Polymers

Associated data

  • ClinicalTrials.gov/NCT02579031