12-Month outcomes of transcatheter tricuspid valve repair with the PASCAL system for severe tricuspid regurgitation

Catheter Cardiovasc Interv. 2021 May 1;97(6):1281-1289. doi: 10.1002/ccd.29583. Epub 2021 Mar 4.

Abstract

Objectives: We investigated the durability of tricuspid regurgitation (TR) reduction and the clinical outcomes through 12 months after transcatheter tricuspid valve repair (TTVr) with the PASCAL Transcatheter Valve Repair System.

Background: TTVr has rapidly developed and demonstrated favorable acute outcomes, but longer follow-up data are needed.

Methods: Overall, 30 patients (age 77 ± 6 years; 57% female) received PASCAL implantation from September 2017 to May 2019 and completed a clinical follow-up at 12 months.

Results: The TR etiology was functional in 25 patients (83%), degenerative in three (10%), and mixed in two (7%). All patients had TR severe or greater (massive or torrential in 80%) and heart failure symptoms (90% in NYHA III or IV) under optimal medical treatment. Single-leaflet device attachment occurred in two patients. Moderate or less TR was achieved in 23/28 patients (82%) at 30 days, which was sustained at 12 months (86%). Two patients underwent repeat TTVr due to residual torrential TR (day 173) and recurrence of severe TR (day 280), respectively. One-year survival rate was 93%; 6 patients required rehospitalization due to acute heart failure. NYHA functional class I or II was achieved in 90% and 6-minute walk distance improved from 275 ± 122 m at baseline to 347 ± 112 m at 12-month (+72 ± 82 m, p < .01). There was no stroke, endocarditis, or device embolization during the follow-up.

Conclusions: Twelve-month outcomes from this multicenter compassionate use experience with the PASCAL System demonstrated high procedural success, acceptable safety, and significant clinical improvement.

Keywords: 12-month outcomes; PASCAL; right-sided heart failure; severe tricuspid regurgitation; transcatheter tricuspid valve intervention.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cardiac Catheterization / adverse effects
  • Female
  • Heart Valve Prosthesis Implantation*
  • Humans
  • Male
  • Time Factors
  • Treatment Outcome
  • Tricuspid Valve / diagnostic imaging
  • Tricuspid Valve / surgery
  • Tricuspid Valve Insufficiency* / diagnostic imaging
  • Tricuspid Valve Insufficiency* / surgery