Vascular complications with a plug-based vascular closure device after transcatheter aortic valve replacement: Predictors and bail-outs

Catheter Cardiovasc Interv. 2021 Nov 1;98(5):E737-E745. doi: 10.1002/ccd.29506. Epub 2021 Feb 3.

Abstract

Background: The MANTA vascular closure device (VCD) is dedicated to large bore access closure and associated with favorable results in selected study populations. Anatomical predictors for access site complications are lacking.

Aim: To evaluate MANTA in a real-world population and identify predictors for vascular complications.

Methods: All patients undergoing transfemoral transcatheter aortic valve replacement (TAVR) between January 2016 and May 2020 with MANTA closure were included. Baseline characteristics were collected, pre-procedural computed tomography and post-deployment femoral angiograms were analyzed for anatomical differences. The primary endpoint was a composite of access site related major and minor vascular complications at 30 days follow-up according to the VARC-2 definitions. Secondary endpoints included bleeding, time to hemostasis, procedural length and incomplete arteriotomy closure or arterial occlusion by angiography. A Cox proportional hazards model was used to compare all-cause mortality for patients with and without an access site complication.

Results: The 512 patients underwent TAVR with MANTA access closure. Median age was 80 (IQR 75-85), 53% was male, median BMI was 26.4 kg/m2 (IQR 23.4-29.7). Access site related major- or minor vascular complication occurred in 20 (4%) and 23 (4%) of patients respectively. Median time to hemostasis was 42 s (IQR 28-98). Post deployment angiogram showed an occlusion in 24 patients (5%), incomplete closure in 60 patients (12%) or both in three patients (1%). Of these 87 patients, 36 (41%) had a vascular complication. Femoral artery diameter (OR 0.70 [0.53-0.93]), low- (OR 3.47 [1.21-10.00]) and high (OR 2.43 [1.16-5.10]) arteriotomies were independent predictors for vascular complications.

Conclusion: In this contemporary TAVR population, access-site related complications occurred in 8% of patients and were mainly due to percutaneous closure device failure. Small artery diameter and off-target punctures were independent predictors.

Keywords: aortic valve disease; mechanical circulatory support; percutaneous intervention; vascular complications.

MeSH terms

  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Catheterization, Peripheral* / adverse effects
  • Female
  • Femoral Artery / diagnostic imaging
  • Femoral Artery / surgery
  • Hemostatic Techniques / adverse effects
  • Humans
  • Male
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Treatment Outcome
  • Vascular Closure Devices*