Long-term intermittent versus short continuous heart rhythm monitoring for the detection of atrial fibrillation recurrences after catheter ablation

Int J Cardiol. 2021 Apr 15:329:105-112. doi: 10.1016/j.ijcard.2020.12.077. Epub 2021 Jan 4.

Abstract

Background: The utility of long-term intermittent heart rhythm monitoring after atrial fibrillation (AF) ablation remains unclear. Therefore, we compared the efficacy and usability of long-term intermittent (AliveCor Kardia® (ACK)) versus short continuous (Holter) heart rhythm monitoring for the detection of AF recurrences after AF ablation and evaluated ACK accuracy to detect AF.

Methods: Patients were provided with Holter (for ≥24 h) simultaneously with an ACK (4 weeks) used three times a day and in case of symptoms. The primary endpoint was the difference in proportion of patients diagnosed with recurrent AF by ACK as compared to Holter monitoring. Secondary endpoints were the usability (System Usability Scale and a four-item questionnaire) of ACK and Holter monitoring; and the accuracy of the ACK algorithm for AF detection.

Results: Out of 126 post-ablation patients, 115 (91.3%; 35 females, median age 64.0 [58.0-68.0] years) transmitted overall 7838 ACK ECG recordings. ACK and Holter monitoring detected 29 (25.2%) and 17 (14.8%) patients with AF recurrences, respectively (p < 0.001). More than 2 weeks of ACK monitoring did not have additional diagnostic yield for detection of AF recurrences. Patients graded ACK higher than Holter monitoring and found ACK more convenient in daily usage than Holter (p < 0.001). Sensitivity and specificity of ACK for AF detection were 95.3% and 97.5%, respectively.

Conclusions: Long-term intermittent monitoring by ACK more effectively detects AF recurrences after AF ablation and has a higher patients' usability than short continuous Holter monitoring. ACK showed a high accuracy to detect AF.

Keywords: AliveCor; Atrial fibrillation; Holter; Telemonitoring; e-health; m-health.

MeSH terms

  • Atrial Fibrillation* / diagnosis
  • Atrial Fibrillation* / surgery
  • Catheter Ablation*
  • Electrocardiography, Ambulatory
  • Female
  • Humans
  • Middle Aged
  • Recurrence
  • Sensitivity and Specificity