Impact of electroanatomical mapping-guided lead implantation on procedural outcome of His bundle pacing

Europace. 2021 Mar 8;23(3):409-420. doi: 10.1093/europace/euaa292.

Abstract

Aims: Conventional His bundle pacing (HBP) can be technically challenging and fluoroscopy-intense, particularly in patients with His-Purkinje conduction disease (HPCD). Three-dimensional electroanatomical mapping (EAM) facilitates non-fluoroscopic lead navigation and HB electrogram mapping. We sought to assess the procedural outcome of routine EAM-guided HBP compared with conventional HBP in a real-world population and evaluate the feasibility and safety of EAM-guided HBP in patients with HPCD.

Methods and results: We included 58 consecutive patients (72 ± 13 years; 71% male) who underwent an attempt to conventional (EAM- group; n = 29) or EAM-guided (EAM+ group; n = 29) HBP between June 2019 and April 2020. The centre's learning curve was initially determined (n = 40 cases) to define the conventional control group and minimize outcome bias favouring EAM-guided HBP. His bundle pacing was successful in 26 patients (90%) in the EAM+ and 27 patients (93%) in the EAM- group (P = 0.64). The procedure time was 90 (73-135) and 110 (70-130) min, respectively (P = 0.89). The total fluoroscopy time [0.7 (0.5-1.4) vs. 3.3 (1.4-6.5) min; P < 0.001] and fluoroscopy dose [21.9 (9.1-47.7) vs. 78.6 (27.2-144.9) cGycm2; P = 0.001] were significantly lower in the EAM+ than EAM- group. There were no significant differences between groups in His capture threshold (1.2 ± 0.6 vs. 1.4 ± 1.0 V/1.0 ms; P = 0.33) and paced QRS duration (113 ± 15 vs. 113 ± 17 ms; P = 0.89). In patients with HPCD, paced QRS duration was similar in both groups (121 ± 15 vs. 123 ± 12 ms; P = 0.77). The bundle branch-block recruitment threshold tended to be lower in the EAM+ than EAM- group (1.3 ± 0.7 vs. 1.8 ± 1.2 V/1.0 ms; P = 0.31). No immediate procedure-related complications occurred. One patient (2%) experienced lead dislodgement during 4-week follow-up.

Conclusion: Implementation of routine EAM-guided HBP lead implantation is feasible and safe in a real-world cohort of patients with and without HPCD and results in a tremendous reduction in radiation exposure without prolonging procedure time or increasing procedure-related complications.

Keywords: Cardiac implantable electronic device; Electroanatomical mapping; His bundle pacing; Implantation technique; Radiation exposure.

MeSH terms

  • Bundle of His*
  • Cardiac Pacing, Artificial
  • Electrocardiography
  • Female
  • Fluoroscopy
  • Humans
  • Male
  • Radiation Exposure*
  • Treatment Outcome