Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study

Europace. 2021 Mar 8;23(3):400-408. doi: 10.1093/europace/euaa307.

Abstract

Aims: Sudden cardiac death (SCD) is the most common mode of death in paediatric hypertrophic cardiomyopathy (HCM). This study describes the implant and programming strategies with clinical outcomes following implantable cardioverter-defibrillator (ICD) insertion in a well-characterized national paediatric HCM cohort.

Methods and results: Data from 90 patients undergoing ICD insertion at a median age 13 (±3.5) for primary (n = 67, 74%) or secondary prevention (n = 23, 26%) were collected from a retrospective, longitudinal multi-centre cohort of children (<16 years) with HCM from the UK. Seventy-six (84%) had an endovascular system [14 (18%) dual coil], 3 (3%) epicardial, and 11 (12%) subcutaneous system. Defibrillation threshold (DFT) testing was performed at implant in 68 (76%). Inadequate DFT in four led to implant adjustment in three patients. Over a median follow-up of 54 months (interquartile range 28-111), 25 (28%) patients had 53 appropriate therapies [ICD shock n = 45, anti-tachycardia pacing (ATP) n = 8], incidence rate 4.7 per 100 patient years (95% CI 2.9-7.6). Eight inappropriate therapies occurred in 7 (8%) patients (ICD shock n = 4, ATP n = 4), incidence rate 1.1/100 patient years (95% CI 0.4-2.5). Three patients (3%) died following arrhythmic events, despite a functioning device. Other device complications were seen in 28 patients (31%), including lead-related complications (n = 15) and infection (n = 10). No clinical, device, or programming characteristics predicted time to inappropriate therapy or lead complication.

Conclusion: In a large national cohort of paediatric HCM patients with an ICD, device and programming strategies varied widely. No particular strategy was associated with inappropriate therapies, missed/delayed therapies, or lead complications.

Keywords: Childhood; Hypertrophic cardiomyopathy; Implantable cardioverter-defibrillator; Programming.

MeSH terms

  • Adolescent
  • Cardiomyopathy, Hypertrophic* / diagnosis
  • Cardiomyopathy, Hypertrophic* / therapy
  • Child
  • Cohort Studies
  • Death, Sudden, Cardiac / epidemiology
  • Death, Sudden, Cardiac / prevention & control
  • Defibrillators, Implantable*
  • Humans
  • Retrospective Studies
  • Risk Factors
  • Treatment Outcome
  • United Kingdom