Reproducibility of global LV function and dyssynchrony parameters derived from phase analysis of gated myocardial perfusion SPECT: A multicenter comparison with core laboratory setting

J Nucl Cardiol. 2022 Jun;29(3):952-961. doi: 10.1007/s12350-020-02397-6. Epub 2020 Oct 20.

Abstract

Background: Gated myocardial perfusion scintigraphy (GMPS) phase analysis is an important tool to investigate the physiology of left ventricular (LV) dyssynchrony. We aimed to test the performance of GMPS LV function and phase analysis in different clinical settings and on a diverse population.

Methods: This is a post hoc analysis of a prospective, non-randomized, multinational, multicenter cohort study. Clinical evaluation and GMPS prior to cardiac resynchronization therapy (CRT)(baseline) and 6-month post CRT (follow-up) were done. LV end-systolic volume (LVESV), LV end-diastolic volume (LVEDV), LV ejection fraction (LVEF), LV phase standard deviation (LVPSD), and percentage of left ventricle non-viable (PLVNV) were obtained by 10 centers and compared to the core lab.

Results: 276 GMPS studies had all data available from individual sites and from core lab. There were no statistically significant differences between all variables except for LVPSD. When subjects with no mechanical dyssynchrony were excluded, LVPSD difference became non-significant. LVESV, LVEF, LVPSD and PLVNV had strong correlation in site against core lab comparison. Bland-Altman plots demonstrated good agreement.

Conclusions: The presented correlation and agreement of LV function and dyssynchrony analysis over different sites with a diverse sample corroborate the strength of GMPS in the management of heart failure in clinical practice.

Keywords: Dyssynchrony; Gated SPECT; Heart failure; MPI; Phase analysis.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cohort Studies
  • Humans
  • Perfusion
  • Prospective Studies
  • Reproducibility of Results
  • Tomography, Emission-Computed, Single-Photon
  • Tomography, X-Ray Computed
  • Ventricular Dysfunction, Left* / diagnostic imaging