Coronary Cannulation After Transcatheter Aortic Valve Replacement: The RE-ACCESS Study

JACC Cardiovasc Interv. 2020 Nov 9;13(21):2542-2555. doi: 10.1016/j.jcin.2020.07.006. Epub 2020 Oct 14.

Abstract

Objectives: The aims of this study were to investigate the feasibility of coronary ostia cannulation after transcatheter aortic valve replacement (TAVR) and to assess potential predictors of coronary access impairment.

Background: Certain data concerning the feasibility and reproducibility of coronary cannulation after TAVR are lacking.

Methods: RE-ACCESS (Reobtain Coronary Ostia Cannulation Beyond Transcatheter Aortic Valve Stent) was an investigator-driven, single-center, prospective, registry-based study that enrolled consecutive patients undergoing TAVR using all commercially available devices. All patients underwent coronary angiography before and after TAVR. The primary endpoint was the rate of unsuccessful coronary ostia cannulation after TAVR. Secondary endpoints were the identification of factors associated with the inability to selectively cannulate coronary ostia after TAVR.

Results: Among 300 patients enrolled in the RE-ACCESS study from December 2018 to January 2020, a total of 23 cases (7.7%) of unsuccessful coronary cannulation after TAVR were documented. This issue occurred in 22 of 23 cases with the use of Evolut R/PRO transcatheter aortic valves (TAVs) (17.9% vs. 0.4%; p < 0.01). In multivariate analysis, the use of Evolut R/PRO TAVs (odds ratio [OR]: 29.6; 95% confidence interval [CI]: 2.6 to 335.0; p < 0.01), the TAV-sinus of Valsalva relation (OR: 1.1 per 1-mm increase; 95% CI: 1.0 to 1.2; p < 0.01), and the mean TAV implantation depth (OR: 1.7 per 1-mm decrease; 95% CI: 1.3 to 2.3; p < 0.01) were found to be independent predictors of unsuccessful coronary cannulation after TAVR. A model combining these factors was demonstrated to predict with very high accuracy the risk for unsuccessful coronary cannulation after TAVR (area under the curve: 0.94; p < 0.01).

Conclusions: Unsuccessful coronary cannulation following TAVR was observed in 7.7% of patients and occurred almost exclusively in those receiving Evolut TAVs. The combination of Evolut TAV, a higher TAV-sinus of Valsalva relation, and implantation depth predicts with high accuracy the risk for unsuccessful coronary cannulation after TAVR. (Reobtain Coronary Ostia Cannulation Beyond Transcatheter Aortic Valve Stent [RE-ACCESS]; NCT04026204).

Keywords: TAVR; aortic stenosis; cannulation; coronary arteries; coronary artery disease.

MeSH terms

  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / surgery
  • Catheterization
  • Heart Valve Prosthesis*
  • Humans
  • Prospective Studies
  • Prosthesis Design
  • Reproducibility of Results
  • Risk Factors
  • Transcatheter Aortic Valve Replacement*
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT04026204