Characteristics of Esophageal Injury in Ablation of Atrial Fibrillation Using a High-Power Short-Duration Setting

Circ Arrhythm Electrophysiol. 2020 Oct;13(10):e008602. doi: 10.1161/CIRCEP.120.008602. Epub 2020 Sep 11.

Abstract

Background: The mechanism of esophageal thermal injury (ETI; esophageal mucosal injury and periesophageal nerve injury leading to gastric hypomotility) remains unknown when using a high-power short-duration (HP-SD) setting. This study sought to evaluate the characteristics of esophageal injuries in atrial fibrillation ablation using a HP-SD setting.

Methods: After exclusion of 5 patients with their esophagus at the right portion of left atrium and 21 patients with additional ablations such as box isolation and low voltage area ablation in left atrium posterior wall, 271 consecutive patients (62±10 years, 56 women) who underwent pulmonary vein isolation by radiofrequency catheter ablation were analyzed. In the 101 patients, a HP-SD setting at 45 to 50 W with an Ablation Index module was used (HP-SD group). In the remaining 170 patients before introduction of the HP-SD setting, a conventional power setting of 20 to 30 W with contact force monitoring was used (conventional group). We performed esophagogastroduodenoscopy after pulmonary vein isolation in all patients and investigated the incidence and characteristics of ETI.

Results: Although the incidence of ETI was significantly higher in the HP-SD group compared with the conventional group (37% versus 22%, P=0.011), the prevalence of esophageal lesions did not differ between the groups (7% versus 8%). Multivariate logistic regression analysis revealed that the use of the HP-SD setting (odds ratio, 6.09, P<0.001), and the parameters that suggest anatomic proximity surrounding the esophagus, were independent predictors of ETI. However, the majority of ETI in the HP-SD group was gastric hypomotility, and the thermal injury was limited to the shallow layer of the periesophageal wall using the HP-SD setting.

Conclusions: Although the use of the HP-SD setting was a strong predictor of ETI, it could avoid deeper thermal injuries that reach the esophageal mucosal layer.

Keywords: atrial fibrillation; catheter ablation; esophagus; incidence; pulmonary vein.

Publication types

  • Comparative Study
  • Observational Study

MeSH terms

  • Aged
  • Atrial Fibrillation / surgery*
  • Burns, Electric / epidemiology
  • Burns, Electric / pathology*
  • Catheter Ablation / adverse effects*
  • Endoscopy, Digestive System
  • Esophagus / injuries
  • Esophagus / pathology*
  • Female
  • Humans
  • Incidence
  • Male
  • Middle Aged
  • Prevalence
  • Pulmonary Veins / surgery*
  • Risk Assessment
  • Risk Factors
  • Treatment Outcome