Individual patient data analysis of the BIOFLOW study program comparing safety and efficacy of a bioresorbable polymer sirolimus eluting stent to a durable polymer everolimus eluting stent

Catheter Cardiovasc Interv. 2021 Nov 1;98(5):848-856. doi: 10.1002/ccd.29254. Epub 2020 Sep 5.

Abstract

Objectives: This analysis of pooled individual patient data (IPD) aimed to evaluate the safety and efficacy of a bioresorbable polymer sirolimus eluting stent system (BP-SES; Orsiro) compared to a durable polymer everolimus eluting stent system (DP-EES; Xience) in the pooled population as well as in subgroups.

Methods: IPD with up to 12 months follow-up of the randomized controlled trials BIOFLOW-II (NCT01356888), -IV (NCT01939249), and -V (NCT02389946) as well as the all comers registry BIOFLOW-III (NCT01553526) were pooled. A total of 3,717 subjects (2,923 in BP-SES and 794 in DP-EES) with 5,328 lesions (4,225 lesions in BP-SES and 1,103 in DP-EES) were included in the IPD. The primary endpoint was target lesion failure (TLF) at 12 months follow-up. Subgroups analyzed included diabetes, age (≥65 years), gender, complex lesions (B2/C), small vessels (reference vessel diameter ≤2.75 mm), multivessel treatment, renal disease, and patients with acute coronary syndrome.

Results: Overall, TLF at 12 months was significantly lower with 5.2%in the BP-SES group versus 7.6% in the DP-EES group (p = .0098). Similarly, target vessel myocardial infarction (TV-MI) was 3.1 versus 5.7% (p = .0005). The rate of stent thrombosis was similar in both groups (0.004%). By regression analysis, an independent stent effect in favor of BP-SES was observed for TLF (p = .0043) and TV-MI (p = .0364) in small vessels.

Conclusion: Results of this IPD analysis suggest that the BP-SES with ultrathin struts is as safe as and more efficacious than DP-EES in the overall cohort and especially in small vessels.

Keywords: drug eluting stent; pooled analysis; small vessel; strut thickness; target lesion failure.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Absorbable Implants
  • Aged
  • Coronary Artery Disease* / diagnostic imaging
  • Coronary Artery Disease* / therapy
  • Data Analysis
  • Drug-Eluting Stents* / adverse effects
  • Everolimus* / adverse effects
  • Humans
  • Polymers
  • Prosthesis Design
  • Sirolimus* / adverse effects
  • Treatment Outcome

Substances

  • Polymers
  • Everolimus
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT01356888
  • ClinicalTrials.gov/NCT01939249
  • ClinicalTrials.gov/NCT02389946
  • ClinicalTrials.gov/NCT01553526