Unexpected high failure rate of a specific MicroPort/LivaNova/Sorin pacing lead

Heart Rhythm. 2021 Jan;18(1):41-49. doi: 10.1016/j.hrthm.2020.08.010. Epub 2020 Aug 13.

Abstract

Background: Pacing leads are the Achilles heel of pacemakers. Most manufacturers report a 3-year survival rate of >99% of their leads. We observed several failures of the Beflex/Vega leads (MicroPort, Shanghai, China; formerly Sorin/LivaNova).

Objective: The purpose of this study was to investigate failure rates of Beflex/Vega leads.

Methods: We analyzed the performance of Beflex/Vega leads implanted at our tertiary referral center. All-cause lead failures (any issues requiring reinterventions such as lead dislocations, cardiac perforations, and electrical abnormalities) were identified during follow-up. The Beflex/Vega lead was compared with a reference lead (CapSureFix Novus 5076, Medtronic, Minneapolis, MN) implanted within the same period and by the same operators.

Results: A total of 585 leads were analyzed (382 Beflex/Vega and 203 CapSureFix Novus 5076 leads). Cumulative failure rate estimates were 5.2%, 6.3%, and 12.4% after 1, 2, and 3 years for the Beflex/Vega lead. This was worse compared to the reference lead (1.5%, 1.5%, 3.7% after 1, 2, and 3 years; P = .001). Early failure manifestations up to 3 months occurred at a similar rate (Beflex/Vega vs CapSureFix Novus 5076 lead: 1.3% vs 0.5% for dislocations; 1.3% vs 1.0% for perforations). During follow-up, electrical abnormalities such as noise oversensing (P = .013) and increased pacing thresholds (P = .003) became more frequent in the Beflex/Vega group. Electrical abnormalities were the most common failure manifestation 3 years after implantation in this group (9.4% vs 2.2% for the CapSureFix Novus 5076).

Conclusion: The failure rate of the Beflex/Vega lead of >10% after 3 years was higher than that of a competitor lead. This gives rise to concern since >135,000 such leads are active worldwide.

Keywords: Beflex; CapSureFix; Complication; Lead failure; Lead survival; Medtronic; MicroPort; Oversensing; Pacemaker; Vega.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Electrodes, Implanted / adverse effects*
  • Equipment Design
  • Equipment Failure / statistics & numerical data*
  • Female
  • Follow-Up Studies
  • Foreign-Body Migration / complications*
  • Foreign-Body Migration / epidemiology
  • Heart Injuries / epidemiology*
  • Heart Injuries / etiology
  • Humans
  • Incidence
  • Male
  • Pacemaker, Artificial / adverse effects*
  • Registries*
  • Retrospective Studies
  • Switzerland / epidemiology