Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T

J Cardiovasc Magn Reson. 2020 May 18;22(1):35. doi: 10.1186/s12968-020-00626-y.

Abstract

Background: Cardiovascular magnetic resonance (CMR) studies in patients with implanted cardioverter/defibrillators (ICD) are increasingly required in daily clinical practice. However, the clinical experience regarding the feasibility as well as clinical value of CMR studies in patients with subcutaneous ICD (S-ICD) is still limited. Besides safety issues, image quality and analysis can be impaired primarily due the presence of image artefacts associated with the generator.

Methods: Twenty-three patients with an implanted S-ICD (EMBLEM, Boston Scientific, Marlborough, Massachusetts, USA; MR-conditional) with suspected cardiomyopathy and/or myocarditis underwent multi-parametric CMR imaging. Studies were performed on a 1.5 T CMR scanner after device interrogation and comprised standard a) balanced steady state free precession cine, b) T2 weighted-edema, c) velocity-encoded cine flow, d) myocardial perfusion, e) late-gadolinium-enhancement (LGE)-imaging and f) 3D-CMR angiography of the aorta. In case of substantial artefacts, alternative CMR techniques such as spoiled gradient-echo cine-sequences and wide-band inversion-recovery LGE (wb-LGE) sequences were applied.

Results: Successful CMR studies could be performed in all patients without any case of unexpected early termination or relevant technical complication other than permanent loss of the S-ICD system beeper volume in 52% of our patients. Assessment of cine-CMR images was predominantly impaired in the left ventricular (LV) anterior, lateral and inferior wall segments and a switch to spoiled gradient echo-based cine-CMR allowed an accurate assessment of cine-images in N = 17 (74%) patients with only limited artefacts. Hyperintensity artefacts in conventional LGE-images were predominantly observed in the LV anterior, lateral and inferior wall segments and image optimisation by use of the wb-LGE was helpful in 15 (65%) cases. Aortic flow measurements and 3D-CMR angiography were assessable in all patients Perfusion imaging artefacts precluded a meaningful assessment in at least one half of the patients. A benefit in clinical-decision making was documented in 17 (74%) patients in the present study.

Conclusion: Safe 1.5 T CMR imaging was possible in all patients with an S-ICD, though the majority had permanent loss of the S-ICD beeper volume. Achieving good image quality may be challenging in some patients - particularly for perfusion imaging. Using spoiled gradient echo-based cine-sequences and wb-LGE sequences may help to reduce the extent of artefacts, thereby allowing accurate cardiac assessment. Thus, 1.5 T CMR studies should not be withhold in patients with S-ICD for safety concerns and/or fear of extensive imaging artefacts precluding successful image analysis.

Keywords: Cardiovascular magnetic resonance; Implanted cardioverter/defibrillator; Late gadolinium enhancement; Wide band.

MeSH terms

  • Adult
  • Aged
  • Artifacts
  • Cardiomyopathies / diagnostic imaging*
  • Cardiomyopathies / physiopathology
  • Cardiomyopathies / therapy
  • Clinical Decision-Making
  • Coronary Circulation
  • Defibrillators, Implantable*
  • Electric Countershock / adverse effects
  • Electric Countershock / instrumentation*
  • Female
  • Humans
  • Magnetic Resonance Imaging, Cine* / adverse effects
  • Male
  • Middle Aged
  • Myocardial Perfusion Imaging / adverse effects
  • Myocardial Perfusion Imaging / methods*
  • Myocarditis / diagnostic imaging*
  • Myocarditis / physiopathology
  • Myocarditis / therapy
  • Patient Safety
  • Predictive Value of Tests
  • Prognosis
  • Prosthesis Design
  • Prosthesis Failure
  • Reproducibility of Results
  • Risk Assessment
  • Risk Factors
  • Ventricular Function, Left