Real-world experience of leadless left ventricular endocardial cardiac resynchronization therapy: A multicenter international registry of the WiSE-CRT pacing system

Heart Rhythm. 2020 Aug;17(8):1291-1297. doi: 10.1016/j.hrthm.2020.03.002. Epub 2020 Mar 9.

Abstract

Background: Biventricular endocardial pacing (BiV ENDO) is a therapy for heart failure patients who cannot receive transvenous epicardial cardiac resynchronization therapy (CRT) or have not responded adequately to CRT. BiV ENDO CRT can be delivered by a new wireless LV ENDO pacing system (WiSE-CRT system; EBR Systems, Sunnyvale, CA), without the requirement for lifelong anticoagulation.

Objective: The purpose of this study was to assess the safety and efficacy of the WiSE-CRT system during real-world clinical use in an international registry.

Methods: Data were prospectively collected from 14 centers implanting the WiSE-CRT system as part of the WiCS-LV Post Market Surveillance Registry. (ClinicalTrials.gov Identifier: NCT02610673).

Results: Ninety patients from 14 European centers underwent implantation with the WiSE-CRT system. Patients were predominantly male, age 68.2 ± 10.5 years, left ventricular ejection fraction 30.6% ± 8.9%, mean QRS duration 180.7 ± 27.0 ms, and 40% with ischemic etiology. Successful implantation and delivery of BiV ENDO pacing was achieved in 94.4% of patients. Acute (<24 hours), 1- to 30-day, and 1- to 6-month complications rates were 4.4%, 18.8%, and 6.7%, respectively. Five deaths (5.6%) occurred within 6 months (3 procedure related). Seventy percent of patients had improvement in heart failure symptoms.

Conclusion: BiV ENDO pacing with the WiSE-CRT system seems to be technically feasible, with a high success rate. Three procedural deaths occurred during the study. Procedural complications mandate adequate operator training and implantation at centers with immediately available cardiothoracic and vascular surgical support.

Keywords: Cardiac resynchronization therapy; Endocardial pacing; Heart failure; Leadless pacing; Nonresponder.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy / methods*
  • Cardiac Resynchronization Therapy Devices*
  • Endocardium
  • Equipment Design
  • Female
  • Heart Failure / physiopathology
  • Heart Failure / therapy*
  • Heart Ventricles / physiopathology*
  • Humans
  • Male
  • Prospective Studies
  • Registries*
  • Stroke Volume / physiology*
  • Treatment Outcome
  • Ventricular Function, Left / physiology*

Associated data

  • ClinicalTrials.gov/NCT02610673