Long-term clinical outcome and performance of transcatheter aortic valve replacement with a self-expandable bioprosthesis

Eur Heart J. 2020 May 21;41(20):1876-1886. doi: 10.1093/eurheartj/ehz925.

Abstract

Aims: In the last decade, transcatheter aortic valve (TAV) replacement determined a paradigm shift in the treatment of patients with severe symptomatic aortic stenosis. Data on long-term TAV performance are still limited. We sought to evaluate the clinical and haemodynamic outcomes of the CoreValve self-expandable valve up to 8-year follow-up (FU).

Methods and results: Nine hundred and ninety inoperable or high-risk patients were treated with the CoreValve TAV in eight Italian Centres from June 2007 to December 2011. The median FU was 4.4 years (interquartile range 1.4-6.7 years). Longest FU reached 11 years. A total of 728 died within 8-year FU (78.3% mortality from Kaplan-Meier curve analysis). A significant functional improvement was observed in the majority of patients and maintained over time, with 79.3% of surviving patients still classified New York Heart Association class ≤ II at 8 years. Echocardiographic data showed that the mean transprosthetic aortic gradient remained substantially unchanged (9 ± 4 mmHg at discharge, 9 ± 5 mmHg at 8 years, P = 0.495). The rate of Grade 0/1 paravalvular leak was consistent during FU with no significant change from post-procedure to FU ≥5 years in paired analysis (P = 0.164). Structural valve deterioration (SVD) and late bioprosthetic valve failure (BVF) were defined according to a modification of the 2017 EAPCI/ESC/EACTS criteria. In cumulative incidence functions at 8 years, moderate and severe SVD were 3.0% [95% confidence interval (CI) 2.1-4.3%] and 1.6% (95% CI 0.6-3.9%), respectively, while late BVF was 2.5% (95% CI 1.2-5%).

Conclusion: While TAVs are questioned about long-term performance and durability, the results of the present research provide reassuring 8-year evidence on the CoreValve first-generation self-expandable bioprosthesis.

Keywords: Aortic stenosis; Durability; Transcatheter aortic valve replacement; Valve deterioration.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / surgery
  • Bioprosthesis*
  • Heart Valve Prosthesis*
  • Humans
  • Prosthesis Design
  • Prosthesis Failure
  • Transcatheter Aortic Valve Replacement*
  • Treatment Outcome