Percutaneous occlusion of transseptal puncture-related free wall perforation at the coronary sinus with a ventricular septal occluder during left atrial appendage closure: A case report

Catheter Cardiovasc Interv. 2020 Dec;96(7):E755-E757. doi: 10.1002/ccd.28589. Epub 2019 Nov 6.

Abstract

Coronary sinus perforation is a life-threatening complication of transseptal puncture and needs to be repaired immediately. In this study, we report a case of a 74-year-old female patient with nonvalvular atrial fibrillation, who is a poor long-term anticoagulation candidate. During the manipulation of transseptal puncture, a perforation of the free right atrial wall at the coronary sinus ostium occurred, which was caused by the Brockenbrough needle and followed by the immediate advancement of an 8.5-French transseptal sheath. In consideration of the danger of cardiac tamponade after sheath removal, we decided to leave the 8.5-French sheath in the pericardial cavity. Then, we advanced a 6 mm ventricular septal occluder through the sheath. Finally, we achieved successful deployment of the device and closure of the perforation under the guidance of fluoroscopy and transthoracic echocardiography. Subsequently, the left atrial appendage orifice was occluded with a 21 mm Watchman device. This case illustrates that percutaneous device closure is feasible for inadvertent perforation of the free right atrial wall at the coronary sinus during transseptal puncture and should be considered as an alternative to surgery.

Keywords: cardiac tamponade; coronary sinus; transseptal puncture.

Publication types

  • Case Reports
  • Video-Audio Media

MeSH terms

  • Aged
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / therapy*
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Coronary Sinus / diagnostic imaging
  • Coronary Sinus / injuries*
  • Female
  • Heart Injuries / diagnostic imaging
  • Heart Injuries / etiology
  • Heart Injuries / therapy*
  • Heart Septum* / diagnostic imaging
  • Hemostatic Techniques / instrumentation*
  • Humans
  • Prosthesis Design
  • Punctures
  • Septal Occluder Device*
  • Treatment Outcome
  • Vascular System Injuries / diagnostic imaging
  • Vascular System Injuries / etiology
  • Vascular System Injuries / therapy*