Safety, efficacy, and hemodynamic performance of a stented bovine pericardial aortic valve bioprosthesis: Two-year analysis

J Thorac Cardiovasc Surg. 2020 Aug;160(2):371-381.e4. doi: 10.1016/j.jtcvs.2019.07.132. Epub 2019 Sep 9.

Abstract

Objectives: The study objectives were to evaluate the safety, efficacy, and hemodynamic performance of a novel stented bovine pericardial aortic valve bioprosthesis 2 years after implantation.

Methods: The PERIcardial SurGical AOrtic Valve ReplacemeNT Pivotal Trial enrolled patients with symptomatic moderate/severe aortic stenosis or regurgitation at centers in Canada, Europe, and the United States. We report the outcomes and hemodynamic performance in patients with up to 2 years of follow-up.

Results: A total of 1273 patients were enrolled, and 1110 underwent implantation. Among patients undergoing implantation, the mean age was 70.2 ± 8.9 years; 833 (75.0%) were male. Risk of mortality (Society of Thoracic Surgeons) was 2.0% ± 1.4%. At the time of analysis, 604 patients had completed the 2-year follow-up visit. Linearized late event rates were as follows: all death, 2.68%; valve-related death, 0.42%; valve thrombosis, 0.05%; endocarditis, 0.94%; thromboembolism, 1.68%; all hemorrhage, 2.94%; major hemorrhage, 1.99%; all paravalvular leak, 0.26%; and major paravalvular leak, 0.05% per patient-year. Mean 2-year aortic gradient and effective orifice area were 13.4 ± 5.0 mm Hg and 1.5 ± 0.37 cm2, respectively. Moderate and severe prosthesis-patient mismatch were observed in 43.5% and 34.8% of patients at 2 years, respectively. Improvement in New York Heart Association class compared with baseline was observed in 73.0% with moderate mismatch and 74.1% with severe mismatch.

Conclusions: The Avalus (Medtronic, Minneapolis, Minn) bovine pericardial valve demonstrates good clinical and safety outcomes at 2 years. Hemodynamic performance shows mean gradients comparable to currently available bovine pericardial aortic valves. There was no clinical impact of moderate to severe mismatch at 2 years. Further follow-up is required to evaluate midterm to long-term clinical outcome.

Keywords: aortic regurgitation; aortic stenosis; bioprosthetic valves; bovine pericardial valves; surgical aortic valve replacement.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Webcast

MeSH terms

  • Aged
  • Animals
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / diagnostic imaging
  • Aortic Valve Insufficiency / mortality
  • Aortic Valve Insufficiency / physiopathology
  • Aortic Valve Insufficiency / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Bioprosthesis*
  • Canada
  • Cattle
  • Europe
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Hemodynamics*
  • Heterografts
  • Humans
  • Male
  • Middle Aged
  • Pericardium / transplantation*
  • Postoperative Complications / etiology
  • Postoperative Complications / mortality
  • Prospective Studies
  • Prosthesis Design
  • Recovery of Function
  • Risk Factors
  • Severity of Illness Index
  • Stents*
  • Time Factors
  • Treatment Outcome
  • United States