The aim of this study was to compare the efficacy and safety of a thin-strut, biodegradable-polymer everolimus-eluting stent (BP-EES) (Synergy, Boston Scientific, Marlborough, Massachusetts) and a thin-strut, durable-polymer everolimus-eluting stent (DP-EES) (XIENCE, Abbott Vascular, Abbott Park, Illinois) in an all-comers population.
Background
BP-EES have been shown to be noninferior to DP-EES in randomized trials in patients at low to moderate risk.
Methods
Among 7,042 consecutive patients who underwent percutaneous coronary intervention between December 2012 and December 2016, 3,870 patients were exclusively treated with BP-EES (n = 1,343) or with DP-EES (n = 2,527). After propensity score matching, the final study population consisted of 1,041 matched patients. The primary endpoint was the device-oriented composite endpoint (cardiac death, target vessel myocardial infarction, and target lesion revascularization) at 12 months.
Results
The device-oriented composite endpoint did not differ between the 2 groups (7.8% with BP-EES vs. 7.1% with DP-EES; hazard ratio: 1.12; 95% confidence interval: 0.81 to 1.53; p = 0.49). There were no differences in rates of cardiac death (3.0% vs. 3.0%; p = 1.00), target vessel myocardial infarction (3.6% vs. 3.1%; p = 0.53), and target lesion revascularization (3.0% vs. 2.5%; p = 0.41). The rate of acute stent thrombosis was significantly higher in the BP-EES group compared with the DP-EES group (1.2% vs. 0.3%; hazard ratio: 4.00; 95% confidence interval: 1.13 to 14.19; p = 0.032). At 12 months, the frequency of definite stent thrombosis did not differ (1.5% vs. 0.9%; hazard ratio: 1.67; 95% confidence interval: 0.73 to 3.82; p = 0.22).
Conclusions
In this consecutively enrolled percutaneous coronary intervention population reflecting routine clinical practice, no difference in the device-oriented composite endpoint between BP-EES and DP-EES was observed throughout 12 months. There was a higher rate of acute stent thrombosis with the BP-EES, a difference that disappeared at 1 year. (CARDIOBASE Bern PCI Registry; NCT02241291)
Central Illustration
Key Words
coronary artery disease
drug-eluting stent
restenosis
stent thrombosis
Abbreviations and Acronyms
BP-EES
biodegradable-polymer everolimus-eluting stent
CI
confidence interval
DES
drug-eluting stent(s)
DOCE
device-oriented composite endpoint
DP-EES
durable-polymer everolimus-eluting stent
HR
hazard ratio
PCI
percutaneous coronary intervention
PS
propensity score(s)
STEMI
ST-segment elevation myocardial infarction
Cited by (0)
Prof. Pilgrim has received research grants to the institution from Biotronik, Symetis/Boston Scientific, and Edwards Lifesciences; and has received speaking fees from Biotronik and Boston Scientific. Prof. Valgimigli has received research grants to the institution from Terumo, Medicure, Abbott, and AstraZeneca; and has received honoraria from Abbott, Chiesi, Bayer, Daiichi-Sankyo, Amgen, Terumo, AstraZeneca, Alvimedica, and Biosensors. Prof. Windecker has received research grants to the institution from Abbott, Amgen, Bayer, Boston Scientific, Biotronik, Edwards Lifesciences, Medtronic, Terumo, and St. Jude Medical. Prof. Räber has received research grants to the institution from Abbott Vascular, Boston Scientific, HeartFlow, Sanofi, and Regeneron; and has received speaking honoraria from Abbott Vascular, Amgen, AstraZeneca, Biotronik, CSL Behring, Sanofi, and Regeneron. All other authors have reported that they have no relationships relevant to the content of this paper to disclose.