Impact of Leaflet Laceration on Transcatheter Aortic Valve-in-Valve Washout: BASILICA to Solve Neosinus and Sinus Stasis

JACC Cardiovasc Interv. 2019 Jul 8;12(13):1229-1237. doi: 10.1016/j.jcin.2019.04.013.

Abstract

Objectives: The aim of this study was to evaluate any potential leaflet washout benefits after bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction during TAVR (BASILICA) in transcatheter valve-in-valve (ViV) in the context of leaflet thrombosis.

Background: Leaflet thrombosis after transcatheter aortic valve replacement is secondary to flow stasis in both the sinus and neosinus. Strategies to improve washout and ameliorate neosinus and sinus flow velocities may have the potential to mitigate the occurrence of clinical and subclinical leaflet thrombosis.

Methods: A 23-mm Edwards SAPIEN 3 and a 26-mm Medtronic Evolut were deployed in a 23-mm transparent surgical aortic valve model before and after leaflet laceration. The valves were placed in the aortic position of a pulse duplicator flow loop. Particle image velocimetry was performed to quantify sinus flow hemodynamic status. A tracing fluorescent dye was injected to evaluate the number of cycles to washout in both regions of interest.

Results: The leaflet laceration procedure led to an increase in the velocities in the sinus and the neosinus by 50% for Evolut ViV and 61.9% for SAPIEN 3 ViV. In addition, leaflet laceration led to a reduction in overall cycles to washout in the neosinus by at least 56% with the Evolut and 54.5% with the SAPIEN 3 and in the sinus by at least 16.7% with the Evolut and 60.8% with the SAPIEN.

Conclusions: Leaflet laceration using a BASILICA-type approach may hold the potential to mitigate neosinus and sinus flow stasis. Controlled in vivo trials are necessary to establish the potential benefit of BASILICA to reduce the occurrence of leaflet thrombosis.

Keywords: BASILICA; TAVR; laceration; leaflet thrombosis; valve-in-valve; washout.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Video-Audio Media

MeSH terms

  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Bioprosthesis*
  • Blood Flow Velocity
  • Coronary Circulation*
  • Coronary Sinus / diagnostic imaging
  • Coronary Sinus / physiopathology*
  • Coronary Stenosis / diagnostic imaging
  • Coronary Stenosis / etiology
  • Coronary Stenosis / physiopathology
  • Coronary Stenosis / prevention & control*
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis*
  • Humans
  • Iatrogenic Disease / prevention & control*
  • Models, Anatomic
  • Prosthesis Design
  • Thrombosis / diagnostic imaging
  • Thrombosis / etiology
  • Thrombosis / physiopathology
  • Thrombosis / surgery*
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Treatment Outcome