Prospective concurrent head-to head comparison of three different types of nitinol occluder device for transcatheter closure of secundum atrial septal defects

EuroIntervention. 2019 Jul 20;15(4):e321-e328. doi: 10.4244/EIJ-D-18-01016.

Abstract

Aims: Previous comparisons between AMPLATZER septal occluders and other designs were retrospective, non-randomised, non-concurrent and involved fewer patients. A prospective concurrent head-to-head comparison of AMPLATZER (ASO), Cera (CSO) and Figulla (FSO) septal occluders was planned to study the patient outcomes.

Methods and results: The three occluders were serially allocated in a cycle of three to consecutively included patients. Demographic, procedural details and complications were analysed. After calculating a sample size of 122 patients in each group, additional patients were recruited to ensure at least 80% follow-up. Four hundred and fifty (450) consecutive patients equally divided among the three designs were comparable in all parameters. There were no major complications and procedural success was 99.6%. The defects and device sizes were similar in all groups; the delivery system was significantly smaller with the ASO. The FSO needed special deployment techniques less often and formed a cobra deformity more often, though this was not statistically significant. Patient outcome was similar among the groups at a follow-up of 12-47 months.

Conclusions: The new occluders are comparable to the ASO with good outcomes and low complication rates in the current era. The new modified structural designs do not show any advantages in terms of procedural complications on early and midterm follow-up, but long-term studies are warranted.

Publication types

  • Comparative Study

MeSH terms

  • Alloys
  • Cardiac Catheterization
  • Heart Septal Defects, Atrial* / surgery
  • Humans
  • Prospective Studies
  • Retrospective Studies
  • Septal Occluder Device*
  • Treatment Outcome

Substances

  • Alloys
  • nitinol