Assessment of left ventricular outflow tract and aortic root: comparison of 2D and 3D transthoracic echocardiography with multidetector computed tomography

Eur Heart J Cardiovasc Imaging. 2019 Oct 1;20(10):1156-1163. doi: 10.1093/ehjci/jez045.

Abstract

Aims: Accurate echocardiographic assessment of left ventricular outflow tract (LVOT) and the aortic root is necessary for risk stratification and choice of appropriate treatment in patients with pathologies of the aortic valve and aortic root. Conventional 2D transthoracic echocardiographic (TTE) assessment is based on the assumption of a circular shaped LVOT and aortic root, although previous studies have indicated a more ellipsoid shape. 3D TTE and multidetector computed tomography (MDCT) applies planimetry and are not dependent on geometrical assumptions. The aim was to test accuracy, feasibility, and reproducibility of 3D TTE compared to 2D TTE assessment of LVOT and aortic root areas, with MDCT as reference.

Methods and results: We examined 51 patients with 2D/3D TTE and MDCT at the same day. All patients were re-examined with 2D/3D TTE on a different day to evaluate 2D and 3D re-test variability. Areas of LVOT, aortic annulus, and sinus were assessed using 2D, 3D TTE, and MDCT. Both 2D/3D TTE underestimated the areas compared to MDCT; however, 3D TTE areas were significantly closer to MDCT-areas. 2D vs. 3D mean MDCT-differences: LVOT 1.61 vs. 1.15 cm2, P = 0.019; aortic annulus 1.96 vs. 1.06 cm2, P < 0.001; aortic sinus 1.66 vs. 1.08 cm2, P = 0.015. Feasibility was 3D 76-79% and 2D 88-90%. LVOT and aortic annulus areas by 3D TTE had lowest variabilities; intraobserver coefficient of variation (CV) 9%, re-test variation CV 18-20%.

Conclusion: Estimation of LVOT and aortic root areas using 3D TTE is feasible, more precise and more accurate than 2D TTE.

Keywords: 2D transthoracic echocardiography; 3D transthoracic echocardiography; Aortic root; Left ventricular outflow tract (LVOT); Multidetector computed tomography (MDCT).

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aortic Diseases / diagnostic imaging*
  • Aortic Diseases / physiopathology
  • Cardiac-Gated Imaging Techniques
  • Contrast Media
  • Denmark
  • Echocardiography / methods*
  • Echocardiography, Three-Dimensional
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Multidetector Computed Tomography*
  • Reproducibility of Results
  • Risk Assessment
  • Triiodobenzoic Acids
  • Ventricular Outflow Obstruction / diagnostic imaging*
  • Ventricular Outflow Obstruction / physiopathology

Substances

  • Contrast Media
  • Triiodobenzoic Acids
  • iodixanol