Predictive value of routine point-of-care cardiac troponin T measurement for prehospital diagnosis and risk-stratification in patients with suspected acute myocardial infarction

Eur Heart J Acute Cardiovasc Care. 2019 Jun;8(4):299-308. doi: 10.1177/2048872617745893. Epub 2017 Dec 4.

Abstract

Objective: The purpose of this study was to determine the predictive value of routine prehospital point-of-care cardiac troponin T measurement for diagnosis and risk stratification of patients with suspected acute myocardial infarction.

Methods and results: All prehospital emergency medical service vehicles in the Central Denmark Region were equipped with a point-of-care cardiac troponin T device (Roche Cobas h232) for routine use in all patients with a suspected acute myocardial infarction. During the study period, 1 June 2012-30 November 2015, prehospital point-of-care cardiac troponin T measurements were performed in a total of 19,615 cases seen by the emergency medical service and 18,712 point-of-care cardiac troponin T measurements in 15,781 individuals were matched with an admission. A final diagnosis of acute myocardial infarction was confirmed in 2187 cases and a total of 2150 point-of-care cardiac troponin T measurements (11.0%) had a value ≥50 ng/l, including 966 with acute myocardial infarction (sensitivity: 44.2%, specificity: 92.8%). Patients presenting with a prehospital point-of-care cardiac troponin T value ≥50 ng/l had a one-year mortality of 24% compared with 4.8% in those with values <50 ng/l, log-rank: p<0.001. The following variables showed the strongest association with mortality in multivariable analysis: point-of-care cardiac troponin T≥50 ng/l (hazard ratio 2.10, 95% confidence interval: 1.90-2.33), congestive heart failure (hazard ratio 1.93, 95% confidence interval: 1.74-2.14), diabetes mellitus (hazard ratio 1.42, 95% confidence interval: 1.27-1.59) and age, one-year increase (hazard ratio 1.08, 95% confidence interval: 1.08-1.09).

Conclusions: Patients with suspected acute myocardial infarction and a prehospital point-of-care cardiac troponin T ≥50 ng/l have a poor prognosis irrespective of the final diagnosis. Routine troponin measurement in the prehospital setting has a high predictive value and can be used to identify high-risk patients even before hospital arrival so that they may be re-routed directly for advanced care at an invasive centre.

Keywords: Acute myocardial infarction; cardiac troponin; emergency medical services; point-of-care; prehospital triage/methods.

Publication types

  • Multicenter Study
  • Observational Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Biomarkers / blood
  • Denmark / epidemiology
  • Electrocardiography
  • Emergency Medical Services / methods*
  • Female
  • Follow-Up Studies
  • Humans
  • Incidence
  • Male
  • Middle Aged
  • Myocardial Infarction / blood
  • Myocardial Infarction / diagnosis*
  • Myocardial Infarction / epidemiology
  • Point-of-Care Systems*
  • Predictive Value of Tests
  • Retrospective Studies
  • Risk Assessment*
  • Survival Rate / trends
  • Time Factors
  • Troponin T / blood*

Substances

  • Biomarkers
  • Troponin T