Quantification of the Effect of Pressure Wire Drift on the Diagnostic Performance of Fractional Flow Reserve, Instantaneous Wave-Free Ratio, and Whole-Cycle Pd/Pa

Circ Cardiovasc Interv. 2016 Apr;9(4):e002988. doi: 10.1161/CIRCINTERVENTIONS.115.002988.

Abstract

Background: Small drifts in intracoronary pressure measurements (±2 mm Hg) can affect stenosis categorization using pressure indices. This has not previously been assessed for fractional flow reserve (FFR), instantaneous wave-free ratio (iFR), and whole-cycle distal pressure/proximal pressure (Pd/Pa) indices.

Methods and results: Four hundred forty-seven stenoses were assessed with FFR, iFR, and whole-cycle Pd/Pa. Cut point values for significance were predefined as ≤0.8, <0.90, and <0.93, respectively. Pressure wire drift was simulated by offsetting the distal coronary pressure trace by ±2 mm Hg. FFR, iFR, and whole-cycle Pd/Pa indices were recalculated and stenosis misclassification quantified. Median (±median absolute deviation) values for FFR, iFR, and whole-cycle Pd/Pa were 0.81 (±0.11), 0.90 (±0.07), and 0.93 (±0.06), respectively. For the cut point of FFR, iFR, and whole-cycle Pd/Pa, 34.6% (155), 50.1% (224), and 62.2% (278) of values, respectively, lay within ±0.05 U. With ±2 mm Hg pressure wire drift, 21% (94), 25% (110), and 33% (148) of the study population were misclassified with FFR, iFR, and whole-cycle Pd/Pa, respectively. Both FFR and iFR had significantly lower misclassification than whole-cycle Pd/Pa (P<0.001). There was no statistically significant difference between the diagnostic performance of FFR and iFR (P=0.125).

Conclusions: In a substantial proportion of cases, small amounts of pressure wire drift are enough to cause stenoses to change classification. Whole-cycle Pd/Pa is more vulnerable to such reclassification than FFR and iFR.

Keywords: diagnosis; diastole; fractional flow reserve, myocardial; physiology; rest.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Artifacts
  • Blood Pressure*
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheters*
  • Coronary Stenosis / classification
  • Coronary Stenosis / diagnosis*
  • Coronary Stenosis / physiopathology
  • Coronary Vessels / physiopathology*
  • Equipment Design
  • Female
  • Fractional Flow Reserve, Myocardial*
  • Humans
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Reproducibility of Results
  • Severity of Illness Index
  • Transducers, Pressure*