Outcomes following transcatheter aortic valve replacement in the United States

JAMA. 2013 Nov 20;310(19):2069-77. doi: 10.1001/jama.2013.282043.

Abstract

Importance: Transcatheter aortic valve replacement (TAVR) was approved by the US Food and Drug Administration for the treatment of severe, symptomatic aortic stenosis and inoperable status (in 2011) and high-risk but operable status (starting in 2012). A national registry (the Society of Thoracic Surgeons/American College of Cardiology Transcatheter Valve Therapy [STS/ACC TVT] Registry) was initiated to meet a condition for Medicare coverage and also facilitates outcome assessment and comparison with other trials and international registries.

Objective: To report the initial US commercial experience with TAVR.

Design, setting, and participants: We obtained results from all eligible US TAVR cases (n=7710) from 224 participating registry hospitals following the Edwards Sapien XT device commercialization (November 2011-May 2013).

Main outcomes and measures: Primary outcomes included all-cause in-hospital mortality and stroke following TAVR. Secondary analyses included procedural complications and outcomes by clinical indication and access site. Device implantation success was defined as successful vascular access, deployment of a single device in the proper anatomic position, appropriate valve function without either moderate or severe AR, and successful retrieval of the delivery system. Thirty-day outcomes are presented for a representative 3133 cases (40.6%) at 114 centers with at least 80% complete follow-up reporting.

Results: The 7710 patients who underwent TAVR included 1559 (20%) cases that were inoperable and 6151 (80%) cases that were high-risk but operable. The median age was 84 years (interquartile range [IQR], 78-88 years); 3783 patients (49%) were women and the median STS predicted risk of mortality was 7% (IQR, 5%-11%). At baseline, 2176 patients (75%) were either not at all satisfied (1297 patients [45%]) or mostly dissatisfied (879 patients [30%]) with their symptom status; 2198 (72%) had a 5-m walk time longer than 6 seconds (slow gait speed). The most common vascular access approach was transfemoral (4972 patients [64%]), followed by transapical (2197 patients [29%]) and other alternative approaches (536 patients [7%]); successful device implantation occurred in 7069 patients (92%; 95% CI, 91%-92%). The observed incidence of in-hospital mortality was 5.5% (95% CI, 5.0%-6.1%). Other major complications included stroke (2.0%; 95% CI, 1.7%-2.4%), dialysis-dependent renal failure (1.9%; 95% CI, 1.6%-2.2%), and major vascular injury (6.4%; 95% CI, 5.8%-6.9%). Median hospital stay was 6 days (IQR, 4-10 days), with 4613 (63%) discharged home. Among patients with available follow-up at 30 days (n=3133), the incidence of mortality was 7.6% (95% CI, 6.7%-8.6%) (noncardiovascular cause, 52%); a stroke had occurred in 2.8% (95% CI, 2.3%-3.5%), new dialysis in 2.5% (95% CI, 2.0%-3.1%), and reintervention in 0.5% (95% CI, 0.3%-0.8%).

Conclusions and relevance: Among patients undergoing TAVR at US centers in the STS/ACC TVT Registry, device implantation success was achieved in 92% of cases, the overall in-hospital mortality rate was 5.5%, and the stroke rate was 2.0%. Although these postmarket US approval findings are comparable with prior published trial data and international experience, long-term follow-up is essential to assess continued efficacy and safety.

Trial registration: clinicaltrials.gov Identifier: NCT01737528.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / surgery
  • Bicuspid Aortic Valve Disease
  • Cardiac Catheterization
  • Cohort Studies
  • Female
  • Femoral Artery
  • Heart Defects, Congenital / surgery*
  • Heart Valve Diseases / surgery*
  • Heart Valve Prosthesis
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis Implantation / mortality*
  • Hospital Mortality*
  • Humans
  • Incidence
  • Length of Stay
  • Male
  • Patient Satisfaction
  • Patient Selection
  • Product Surveillance, Postmarketing
  • Registries / statistics & numerical data
  • Stroke / epidemiology*
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01737528