Cardiac tamponade as an inherent but potentially nonfatal complication of transvenous lead extraction: Experience with 1126 procedures performed using mechanical tools

J Cardiovasc Electrophysiol. 2022 Dec;33(12):2625-2639. doi: 10.1111/jce.15668. Epub 2022 Sep 14.

Abstract

Introduction: Cardiac tamponade (CT) is one of the most common and dangerous complications of transvenous lead extraction (TLE). So far, however, there has been little discussion about the problem.

Methods: We analyzed the occurrence of CT in a group of 1226 patients undergoing TLE at a single reference center between June, 2015 and February, 2021. Using standard mechanical devices as first-line tools, a total of 2092 leads had been extracted.

Results: CT occurred in 18 patients (1.47%): due to injury to the wall of the right atrium in 14 patients (1.14%) and other cardiac walls in four patients (0.33%). Younger patient age at first implantation, female gender, high left ventricular ejection fraction (LVEF), lower New York Heart Association class, low Charlson comorbidity index, longer implant duration, and the number of previous procedures related to cardiac implantable electronic devices (CIED) are important patient-related risk factors for CT. Significant procedure-related risk factors include the number of extracted leads, extraction of atrial leads and longer dwell time of extracted leads. Intraoperative transoesophageal echocardiography (TEE) provides a lot of information about pulling on various cardiac structures and is able to detect a very early phase of bleeding to the pericardial sac. As a result of implementing best practices guidance in performing extraction procedures and close collaboration with cardiac surgeons that allowed immediate rescue intervention in our series of 18 CT cases, there were no procedure-related deaths (mortality 0%).

Conclusions: The need for rescue surgery due to CT has no influence on clinical and procedural success. Early diagnosed (TEE monitoring) and properly managed CT does not generate any additional risk in short- and long-term follow-up after TLE.

Keywords: cardiac tamponade; lead extraction complications; pericardial fluid; surgery in lead extraction; transvenous lead extraction.

MeSH terms

  • Cardiac Tamponade* / diagnostic imaging
  • Cardiac Tamponade* / etiology
  • Cardiac Tamponade* / therapy
  • Defibrillators, Implantable* / adverse effects
  • Device Removal / adverse effects
  • Device Removal / methods
  • Equipment Failure
  • Female
  • Humans
  • Pacemaker, Artificial* / adverse effects
  • Retrospective Studies
  • Stroke Volume
  • Treatment Outcome
  • Ventricular Function, Left