Hemodynamic response and safety of vasodilator stress cardiovascular magnetic resonance in patients with permanent pacemakers or implantable cardioverter-defibrillators

J Cardiovasc Electrophysiol. 2022 Oct;33(10):2127-2135. doi: 10.1111/jce.15630. Epub 2022 Jul 28.

Abstract

Introduction: Vasodilator stress cardiovascular magnetic resonance (CMR) is a powerful diagnostic modality, but data toward its use in patients with permanent pacemakers (PPMs) or implantable cardioverter-defibrillators (ICDs) is limited.

Methods and results: Patients with ICDs (>1% pacing) or PPMs who underwent regadenoson single photon emission computed tomography (SPECT) and all patients with ICDs or PPMs who underwent stress CMR were retrospectively identified. SPECT tests were analyzed for hemodynamic responses and new pacing requirements; CMR studies were examined for safety, device characteristics and programming, hemodynamic responses, and image quality. Changes from baseline were evaluated with the Related-Samples Wilcoxon Signed Rank Test. Of 67 patients (median age 65 [IQR 58-72] years, 31 [46%] female, 31 [46%] Black), 47 underwent SPECT and 20 CMR. With regadenoson SPECT, 89% of patients experienced tachycardic responses above resting heart rates (+19 [13-32] beats per minute, p < .01). During stress CMR, 10 (50%) devices were asynchronously paced approximately 10 beats per minute above resting rates, and the remaining were temporarily deactivated. Those with asynchronous pacing had no changes in heart rates, whereas patients with deactivated devices had near uniform heart rate accelerations. Image quality was diagnostic in the majority of stress CMR sequences, with nonconditional ICDs contributing 40 of 57 (70%) of nondiagnostic segments.

Conclusion: This data supports the safety of vasodilator stress CMR with promising diagnostic quality images in patients with CMR conditional ICDs and PPMs. Despite a near uniform tachycardic response to regadenoson in the SPECT environment, high rates of asynchronous pacing during vasodilator stress CMR did not result in competitive pacing or adverse arrhythmic events. Further studies are needed to validate these findings and confirm the diagnostic and prognostic performance of stress CMR in these individuals.

Keywords: artificial cardiac pacemaker; cardioverter defibrillator; implantable; magnetic resonance imaging; single photon emission computed tomography; stress test.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Aged
  • Defibrillators, Implantable* / adverse effects
  • Female
  • Heart Rate
  • Humans
  • Magnetic Resonance Imaging
  • Magnetic Resonance Spectroscopy
  • Male
  • Pacemaker, Artificial* / adverse effects
  • Retrospective Studies
  • Vasodilator Agents