New Research Paper
Structural
Balloon-Expandable Valve for Treatment of Evolut Valve Failure: Implications on Neoskirt Height and Leaflet Overhang

https://doi.org/10.1016/j.jcin.2021.12.021Get rights and content
Under an Elsevier user license
open archive

Abstract

Objectives

This study sought to determine the degree of Evolut (Medtronic) leaflet pinning, diameter expansion, leaflet overhang, and performance at different implant depths of the balloon-expandable Sapien 3 (S3, Edwards Lifesciences LLC) transcatheter heart valve (THV) within the Evolut THV.

Background

Preservation of coronary access and flow is a major factor when considering the treatment of failed Evolut THVs.

Methods

An in vitro study was performed with 20-, 23-, 26-, and 29-mm S3 THVs deployed within 23-, 26-, 29-, and 34-mm Evolut R THVs, respectively. The S3 outflow was positioned at various depths at node 4, 5, and 6 of the Evolut R. Neoskirt height, leaflet overhang, performance, and Evolut R valve housing diameter expansion were assessed under physiological conditions as per ISO 5840-3 standard.

Results

The neoskirt height for the Evolut R was shorter when the S3 outflow was positioned at node 4 compared with node 6 (node 4 height for 23 mm = 16.3 mm, 26 mm = 17.1 mm, 29 mm = 18.3 mm, and 34 mm = 19.9 mm vs node 6 height for 23 mm = 23.9 mm, 26 mm = 23.4 mm, 29 mm = 24.7 mm, and 34 mm = 27 mm Evolut R). All configurations exhibited acceptable hydrodynamic performance irrespective of the degree of leaflet overhang, except the 29-mm S3 implanted in 34-mm Evolut R at node 4 (regurgitant fraction >20%). The valve housing radius of the index Evolut R increased when the S3 was implanted, with the increase ranging from 0 to 2.5 mm.

Conclusions

Placement of the S3 at a lower implant position within an index Evolut R reduces the neoskirt height with no significant compromise to S3 valve function despite a higher degree of leaflet overhang. Low S3 implantation may facilitate future coronary access after redo transcatheter aortic valve replacement.

Key Words

aortic stenosis
bench study
redo transcatheter aortic valve replacement

Abbreviations and Acronyms

S3
Sapien 3
TAVR
transcatheter aortic valve replacement
THV
transcatheter heart valve

Cited by (0)

The authors attest they are in compliance with human studies committees and animal welfare regulations of the authors’ institutions and Food and Drug Administration guidelines, including patient consent where appropriate. For more information, visit the Author Center.