"But for the blind spot": Accuracy and diagnostic performance of smart watch cardiac features in pediatric patients

Heart Rhythm. 2024 May;21(5):581-589. doi: 10.1016/j.hrthm.2024.01.021. Epub 2024 Jan 20.

Abstract

Background: The Apple Watch™ (AW) offers heart rate (HR) tracking by photoplethysmography (PPG) and single-lead electrocardiographic (ECG) recordings. The accuracy of AW-HR and diagnostic performance of AW-ECGs among children during both sinus rhythm and arrhythmias have not been explored.

Objective: The purposes of this study were to assess the accuracy of AW-HR measurements compared to gold standard modalities in children during sinus rhythm and arrhythmias and to identify non-sinus rhythms using AW-ECGs.

Methods: Subjects ≤18 years wore an AW during (1) telemetry admission, (2) electrophysiological study (EPS), or (3) exercise stress test (EST). AW-HRs were compared to gold standard modality values. Recorded AW-ECGs were reviewed by 3 blinded pediatric electrophysiologists.

Results: Eighty subjects (median age 13 years; interquartile range 1.0-16.0 years; 50% female) wore AW (telemetry 41% [n = 33]; EPS 34% [n = 27]; EST 25% [n = 20]). A total of 1090 AW-HR measurements were compared to time-synchronized gold standard modality HR values. Intraclass correlation coefficient (ICC) was high 0.99 (0.98-0.99) for AW-HR during sinus rhythm compared to gold standard modalities. ICC was poor comparing AW-HR to gold standard modality HR in tachyarrhythmias (ICC 0.24-0.27) due to systematic undercounting of AW-HR values. A total of 126 AW-ECGs were reviewed. Identification of non-sinus rhythm by AW-ECG showed sensitivity of 89%-96% and specificity of 78%-87%.

Conclusions: We found high levels of agreement for AW-HR values with gold standard modalities during sinus rhythm and poor agreement during tachyarrhythmias, likely due to hemodynamic effects of tachyarrhythmias on PPG-based measurements. AW-ECGs had good sensitivity and moderate specificity in identification of non-sinus rhythm in children.

Keywords: Apple Watch; Arrhythmia; Pediatric patient; Photoplethysmography; Wearable technology.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Arrhythmias, Cardiac* / diagnosis
  • Arrhythmias, Cardiac* / physiopathology
  • Child
  • Child, Preschool
  • Electrocardiography / methods
  • Electrocardiography, Ambulatory / instrumentation
  • Electrocardiography, Ambulatory / methods
  • Equipment Design
  • Exercise Test / methods
  • Female
  • Heart Rate* / physiology
  • Humans
  • Infant
  • Male
  • Photoplethysmography* / instrumentation
  • Photoplethysmography* / methods
  • Reproducibility of Results
  • Telemetry / instrumentation
  • Telemetry / methods
  • Wearable Electronic Devices