Use of the Inspiris valve in the native right ventricular outflow tract is associated with early prosthetic regurgitation

J Thorac Cardiovasc Surg. 2023 Oct;166(4):1210-1221.e8. doi: 10.1016/j.jtcvs.2023.04.018. Epub 2023 Apr 22.

Abstract

Objective: The Inspiris Resilia prosthesis (Edwards Lifesciences) has been increasingly used in the pulmonic position with limited performance data. We sought to investigate its durability as a surgical pulmonary valve replacement (PVR).

Methods: We retrospectively reviewed patients who underwent PVR or conduit replacement with an Inspiris or non-Inspiris valve/conduit from 2018 to 2022. The primary end point was freedom from a composite of at least moderate pulmonary regurgitation, pulmonary stenosis, or valve/conduit reintervention. Secondary end points were individual components of the composite outcome. To account for baseline differences, propensity matching identified 70 patient pairs.

Results: A total of 227 patients (median age: 19.3 years [interquartile range, 11.8-34.4]) underwent PVR or conduit replacement (Inspiris: n = 120 [52.9%], non-Inspiris: n = 107 [47.1%]). Median follow-up was 26.6 months [interquartile range, 12.4-41.1]. Among matched patients, 2-year freedom from valve failure was lower in the Inspiris group (53.5 ± 9.3% vs 78.5 ± 5.9%, P = .03), as was freedom from at least moderate pulmonary regurgitation (54.2 ± 9.6% vs 86.4 ± 4.9%, P < .01). There was no difference in 2-year freedom from at least moderate pulmonary stenosis (P = .61) or reintervention (P = .92). Inspiris durability was poorer when implanted in the native right ventricular outflow tract compared with as a conduit, with 18-month freedom from valve failure of 59.0 ± 9.5% versus 85.9 ± 9.5% (P = .03).

Conclusions: Early durability of the Inspiris valve is poor when implanted in the native right ventricular outflow tract; its unique design may be incompatible with the compliant pulmonary root. Modified implantation techniques or alternative prostheses should be considered.

Keywords: bioprosthetic valves; pulmonary valve replacement; right ventricular outflow tract; transcatheter pulmonary valve implantation.

MeSH terms

  • Adult
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Heart Valve Prosthesis Implantation* / methods
  • Heart Valve Prosthesis*
  • Humans
  • Prosthesis Design
  • Pulmonary Valve Insufficiency* / diagnostic imaging
  • Pulmonary Valve Insufficiency* / etiology
  • Pulmonary Valve Insufficiency* / surgery
  • Pulmonary Valve Stenosis* / diagnostic imaging
  • Pulmonary Valve Stenosis* / surgery
  • Pulmonary Valve* / diagnostic imaging
  • Pulmonary Valve* / surgery
  • Retrospective Studies
  • Treatment Outcome
  • Ventricular Outflow Obstruction* / diagnostic imaging
  • Ventricular Outflow Obstruction* / etiology
  • Ventricular Outflow Obstruction* / surgery
  • Young Adult