The efficacy and safety of Gore conformable thoracic stent graft and Valiant Captivia thoracic stent graft for acute type B aortic dissection

Int J Cardiol. 2023 Jul 1:382:3-11. doi: 10.1016/j.ijcard.2023.03.060. Epub 2023 Apr 3.

Abstract

Background: To evaluate the safety and efficacy of the conformable thoracic aortic endograft (Conformable TAG Thoracic Endoprosthesis [CTAG]; W. L. Gore & Associates, Flagstaff, Ariz) and Valiant Captivia thoracic stent graft (Medtronic Inc., Santa Rosa, CA) for acute type B aortic dissection (TBAD).

Methods: The early and mid-term outcomes were analyzed for 413 patients undergoing TEVAR using conformable TAG thoracic endoprosthesis and Valiant Captivia thoracic stent graft for acute TBAD. 100 propensity-matched pairs of patients were generated, including 100 patients in the CTAG group and 200 patients in the Valiant Captivia group.

Results: Operative mortality were 2.33% (3 of 129) in the CTAG group and 1.76% (5 of 284) in the Valiant Captivia group. The median follow-up was 41.67 (26.00-60.67) months. No significant difference in mortality (9 [7.00%] vs. 36 [12.68%], P = 0.95) or re-intervention rate (3 [2.33%] vs. 20 [7.04%], P = 0.29) was observed between two groups. CTAG group have a lower incidence rate of distal stent graft-induced new entry tear than Valiant Captivia group (2.33% vs. 9.86%, P = 0.045). Lower incidence of type Ia endoleak was identified in the CTAG group (2.22%) than the Valiant Captivia group (14.41%) in patients with type III arch (P = 0.039).

Conclusions: Both Valiant Captivia thoracic stent graft and CTAG thoracic endoprosthesis can be safely performed for acute TBAD with low operative mortality, favorable mid-term survival and freedom from reintervention. CTAG thoracic endoprosthesis had fewer dSINE even with larger oversizing and potentially suitable for type III arch with fewer type Ia endoleaks.

Keywords: Conformable thoracic endograft; Distal stent graft-induced new entry tear; I endoleaks; Thoracic endovascular aortic repair; Type B aortic dissection.

MeSH terms

  • Aortic Aneurysm, Thoracic* / complications
  • Aortic Aneurysm, Thoracic* / diagnostic imaging
  • Aortic Aneurysm, Thoracic* / surgery
  • Aortic Dissection* / diagnostic imaging
  • Aortic Dissection* / surgery
  • Blood Vessel Prosthesis / adverse effects
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Endoleak
  • Endovascular Procedures* / adverse effects
  • Humans
  • Prosthesis Design
  • Stents / adverse effects
  • Time Factors
  • Treatment Outcome