First-in-Human Dedicated Leaflet Splitting Device for Prevention of Coronary Obstruction in Transcatheter Aortic Valve Replacement

JACC Cardiovasc Interv. 2023 Jan 9;16(1):94-102. doi: 10.1016/j.jcin.2022.10.050.

Abstract

Background: Coronary artery obstruction is a life-threatening complication of transcatheter aortic valve replacement (TAVR) procedures. Current preventive strategies are suboptimal.

Objectives: The aim of this study was to describe bench testing and clinical experience with a novel device that splits valve leaflets that are at risk for causing coronary obstruction after TAVR, allowing normal coronary flow.

Methods: The ShortCut device was initially tested in vitro and preclinically in a porcine model for functionality and safety. The device was subsequently offered to patients at elevated risk for coronary obstruction. Risk for coronary obstruction was based on computed tomography-based anatomical characteristics. Procedure success was determined as patient survival at 30 days with a functioning new valve, without stroke or coronary obstruction.

Results: Following a successful completion of bench testing and preclinical trial, the device was used in 8 patients with failed bioprosthetic valves (median age 81 years; IQR: 72-85 years; 37.5% man) at 2 medical centers. A total of 11 leaflets were split: 5 patients (63.5%) were considered at risk for left main obstruction alone, and 3 patients (37.5%) were at risk for double coronary obstruction. All patients underwent successful TAVR without evidence of coronary obstruction. All patients were discharged from the hospital in good clinical condition, and no adverse neurologic events were noted. Procedure success was 100%.

Conclusions: Evaluation of the first dedicated transcatheter leaflet-splitting device shows that the device can successfully split degenerated bioprosthetic valve leaflets. The procedure was safe and successfully prevented coronary obstruction in patients at risk for this complication following TAVR.

Keywords: coronary artery obstruction; structural heart disease; transcatheter aortic valve replacement; valve-in-valve.

MeSH terms

  • Aged, 80 and over
  • Animals
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / complications
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Bioprosthesis* / adverse effects
  • Coronary Occlusion* / diagnostic imaging
  • Coronary Occlusion* / etiology
  • Coronary Occlusion* / prevention & control
  • Heart Valve Prosthesis* / adverse effects
  • Humans
  • Male
  • Prosthesis Design
  • Prosthesis Failure
  • Risk Factors
  • Swine
  • Transcatheter Aortic Valve Replacement*
  • Treatment Outcome