One-year clinical and echocardiographic outcomes of transcatheter aortic valve replacement with SAPIEN 3 Ultra versus SAPIEN 3

Catheter Cardiovasc Interv. 2022 Dec 13. doi: 10.1002/ccd.30515. Online ahead of print.

Abstract

Background: The newest-generation balloon-expandable valve, SAPIEN 3 Ultra (S3U), is expected to function well with the enhanced sealing skirt compared with SAPIEN 3 (S3). However, current literature on the comparison between these two valves is limited to short-term follow-ups. Therefore, we aimed to evaluate the 1-year outcomes and echocardiographic changes of S3U compared with S3.

Methods: We retrospectively identified patients who underwent transfemoral-transcatheter aortic valve replacement with S3U or S3 (20/23/26 mm) at our institution in 2018-2020. Outcomes were 1-year clinical events and echocardiographic parameters, and were compared between S3U and S3 after adjustment with inverse probability of treatment weighting.

Results: The S3U and S3 groups included 297 (25.7%) and 858 (74.3%) patients, respectively. There were no significant differences between the two groups in clinical events (death 5.8% vs. 5.5%, hazard ratio [HR] 1.07, 95% confidence interval [CI] 0.81-1.90; death or heart failure hospitalization 10.3% vs. 10.1%, HR 1.04, 95% CI 0.67-1.62). The S3U group had a lower prevalence of mild paravalvular leakage (PVL) (13.7% vs. 22.3%, p = 0.044), with similar moderate PVL (0.7% vs. 1.2%, p > 0.99). No significant differences were observed in aortic valve mean gradient and Doppler velocity index at 1 year. However, the S3U group had a larger increase in mean gradient from implantation to 1 year (median +4.70 vs. +1.63 mmHg, p < 0.001).

Conclusions: S3U and S3 carried similarly favorable clinical event risks. Nonetheless, S3U was associated with less frequent mild PVL but a larger increase in transprosthetic gradient. Further studies are needed to determine the prognostic impact of these hemodynamic differences.

Keywords: SAPIEN 3 Ultra; balloon-expandable valve; hemodynamics; paravalvular leakage; transcatheter aortic valve replacement.

Grants and funding