Effects of different ablation settings on lesion dimensions in an ex vivo swine heart model: Baseline impedance, irrigant, and electrode configuration

J Cardiovasc Electrophysiol. 2023 Jan;34(1):117-125. doi: 10.1111/jce.15752. Epub 2022 Nov 30.

Abstract

Introduction: Intramural or epicardial locations of the arrhythmogenic substrate are regarded as one of the main reasons for radiofrequency (RF) catheter ablation failure. This study aims to conduct a comprehensive analysis of various factors including baseline impedance, irrigant and electrode configuration at similar ablation index (AI) value.

Methods: In 12 ex vivo swine hearts, RF ablation was performed at a target AI value of 500 and a multistep impedance load (100-180 Ω) in 4 settings: (1) conventional unipolar configuration with an irrigant of normal saline (NS); (2) conventional unipolar configuration with an irrigant of half normal saline (HNS); (3) bipolar configuration with an irrigant of NS; (4) sequential unipolar configuration with an irrigant of NS. The relationships between lesion dimensions and above factors were examined.

Results: Baseline impedance had a strong negative linear correlation with lesion dimensions at a certain AI. The correlation coefficient between baseline impedance and depth, width, and volume were R = -0.890, R = -0.755 and R = -0.813, respectively (p < .01). There were 10 (total: 10/100, 10%; bipolar: 10/25, 40%) transmural lesions during the whole procedure. Bipolar ablation resulted in significantly deeper lesion than other electrode configurations. Other comparisons in our experiment did not achieve statistical significance.

Conclusion: There is a strong negative linear correlation between baseline impedance and lesion dimensions at a certain AI value. Baseline impedance has an influence on the overall lesion dimensions among irrigated fluid and ablation configurations. Over a threshold impedance of 150 Ω, the predictive accuracy of AI can be compromised.

Keywords: HNS; baseline impedance; bipolar ablation; lesion dimensions; radiofrequency catheter ablation; sequential unipolar ablation.